Accuracy is guaranteed for one year at 23±5°C (73±9°F) and Max. 80% RH. (Opening and Closing of the Sensor: Maximum 10000 times) |
Rated current |
10 A AC |
Output voltage |
100 mV AC/A |
Amplitude accuracy |
±1.0% rdg.±0.05% f.s. (f.s.: 10A) (at the clamp core center, 45 to 66 Hz) |
Phase accuracy |
within ±3°(at 50Hz, 60Hz, Measurement current: Min. 1A) |
Maximum input current |
30A continuous at 45 to 60Hz (Ambient temperature: 50°C) |
Effect of conductor position |
within ±0.1% (in any direction from sensor center) |
Effect of external magnetic fields |
400 A AC/m corresponds to 5 mA, Max. 7.5 mA |
Remaining electric current character |
Max. 5 mA (in 100 A go and return electric wire) |
Dielectric strength |
3536 Vrms for 15 seconds (between clamp window and case) |
Maximum rated voltage to earth |
Max. 300 Vrms (Insulated conductor) |
Operating temperature and humidity range |
0 to 50°C (32 to 122°F) Max. 80%RH (no condensation) |
Storage temperature and humidity range |
-10 to 60°C (14 to 140°F) Max. 80%RH (no condensation) |
Location for use |
Altitude up to 2000 m (6562 feet), Indoors |
Standards applying |
EMC
EN 61326:1997+A1:1998+A2:2001+A3:2003
Safety
EN 61010-2-032:2002 Type B current sensor Measurement Category III, Pollution Degree 2 (anticipated transient overvoltage 4000 V) |
Diameter of measurable conductor |
Within Φ 40 mm (1.57") |
Cable length |
Approx. 3 m (118.11") |
External dimensions |
Approx. 74W x 145H x 42D mm 2.91"W x 5.71"H x 1.65"D (excluding protrusions) |
Mass |
Approx. 380 g (12.2 oz.) |
Accessories |
Instruction manual: 1 |
We define measurement tolerances in terms of f.s. (full scale) and rdg.(reading) values, with the following meanings:
f.s. (maximum display value or scale length) The maximum displayable value or the full length of the scale. This is usually the maximum value of the currently selected range.
rdg. (reading or displayed value) The value currently being measured and indicated on the measuring product. |