Models |
ST5540 |
ST5541 |
Measurement methods |
Measurement of voltage drop across body simulated resistance points
Calculation and display of current values True rms measurement
Measurement unit floats relative to instrument ground. |
Measurement modes |
Leak current measurement, voltage measurement, safety conductor current measurement |
Leak current measurement |
Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current |
Measurement current |
DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz) |
Measurement ranges |
DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA
AC peak mode:500 μA/1 mA/10 mA/75 mA |
Measurement accuracy (current measurement) |
DC measurement: ?2.0% rdg. ?6 dgt. (typ.)
AC / AC+DC measurement: ?2.0% rdg. ?6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ?2.0% rdg. ?6 dgt. (15 Hz to 10 kHz, typ.) |
Interfaces |
External I/O, medical device relay output, USB 1.1 (communications), RS-232C |
Functionality |
110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc. |
Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc. |
Power supply |
100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power |
Target device power supply input |
100 to 240 V AC, 50/60 Hz Rated current input from terminal block: 20 A |
Target device power supply output |
Output from terminal block: 20 A Output from outlet: 15 A |
Dimensions, mass |
320 (W) ? 110 (H) ? 253 (D) mm, 4.5 kg |
Accessories |
1 - Test Lead 9170-10
1 - Enclosure probe 9195
1 - Test Lead Red
3 - Alligator Clips (1 red and 1 black)
1 - Instruction manual
1 - Users manual
3 - Power cord
1 - Spare measurement fuse |
1 - Test Lead 9170-10
1 - Enclosure probe 9195
3 - Alligator Clips (1 red and 1 black)
1 - Instruction manual
1 - Users manual
3 - Power cord
1 - Spare measurement fuse |