Yokogawa BT200 STANDARD SPECIFICATIONS |
Applicable Equipment Remove |
DPharp
ADMAG
YEWFLO
YTA
Signal Conditioner |
Communication Signal Connection |
Dedicated cable, 1.1 m long (3.6 ft) |
Communication Line |
Line length |
Up to 2 km (1.24 mile) (0.75 to 1.25 mm 2 instrumentation cable) |
Load resistance |
250 to 600 Ω (including cable resistance) |
Load capacitance |
0.22 μ F or less |
Load inductance |
3.3 mH or less |
Power line spacing |
15 cm (6 inch) or more (Avoid parallel wiring.) |
EMC Conformity Standards |
EN61326-1 Class A, Table2 (For use in industrial locations), EN61326-2-3 |
Display |
LCD dot matrix, 21 characters X 8 lines |
Controls |
Membrane switches (four function keys, 20 general operation keys, and one power switch) |
Printer (BT200-P00) |
Thermal paper type |
Power Supply |
Five AA 1.5 V dry alkali batteries (LR6/AM3 (N)) For Intrinsic safety type; Five designated batteries |
Dimensions |
BT200-N00 |
228 X110 X 51 mm
(9.0 X 4.3 X 2.0 inch) |
BT200-P00 |
321 X 110 X 61 mm
(12.6 X 4.3 X 2.4 inch) |
Approximate Weight |
BT200-N00 |
510 g (1.12 lb) |
BT200-P00 |
700 g (1.54 lb) |
Functional Specifications |
Basic Functions |
Setup, alteration, and display of parameters BRAIN communication. |
constant current output |
Zero point adjustment |
Additional Functions |
Batch upload/download of data
Setpoint Protection: Security code entry is required to alter setpoints.
Battery Alarm: An alarm message appearing on the LCD announces low battery voltages.
Automatic Power-off: The terminal is switched off automatically if no key access is made for approximately 5 minutes.
LCD contrast adjustment
Printing (BT200-P00) |
Printout Information |
All parameter lists
Parameter list within each menu item
Setup change data list
Uploaded data list
Display images
Self check list |
NORMAL OPERATION CONDITIONS |
BT200-N00 |
Ambient Temperature |
-15 to 55 °C (5 to 131 °F) |
Ambient Humidity |
5 to 95%RH at 40 °C (104 °F) |
BT200-P00 |
Ambient Temperature |
0 to 50 °C (32 to 122 °F) |
Ambient Humidity |
30 to 80%RH at 40 °C (104 °F) |
STORAGE CONDITIONS |
Ambient Temperature |
-15 to 60 °C (5 to 140 °F) |
Ambient Humidity |
5 to 95%RH at 40 °C (104 °F)
30 to 80%RH at 40 °C (104 °F) (with printer) |