YSI IDS 4210W
ORP Wireless Sensor
|
- Combination of ORP
- Glass Material
- Refillable (3 mol KCl)
- Digital
- With Double junction probe
- Wireless Sensor
|
The IDS sensors automatically store their unique serial number and calibration data. In addition, they also digitally process the measurement signal. The IDS sensors can be moved from instrument to instrument and maintain their calibration data and transmit that information to the new instrument.
One WA-S wireless sensor module must be installed at the top of each wireless sensor, while one WA-M wireless meter module is needed per MultiLab meter. We suggest the MultiLab WA Kit if wireless sensors will be used, as it includes one WA-S wireless sensor module, one WA-M wireless meter module, one charger for WA-S modules, and one USB power supply for the WA-S charger.
Wireless sensors can only be used with the MultiLab 4010-1W, MultiLab 4010-2W, and the MultiLab 4010-3W. Wireless sensors can be used with the MultiLab 4010-1 if the latest instrument firmware has been installed - see the Software and Firmware Downloads page.
- If necessary, clean the plug connection.
- Open the locking device (step 1)
- Use your thumb and index finger to press the clips of the connector together, and pull the connector out of the plug (step 2)
- Make sure that the plug connection is completely dry and clean.
- Align the guidance of the connector with the groove in the plug and insert the connector in the unlocked plug until it catches (step 3)
- Close the locking device (step 4)