RF Input |
Frequency Range |
1.5MHz to 2.0GHz with a useful response, with reduced sensitivity, to at least 4GHz. |
Tuning |
Automatic tuning selects the largest available signal. Correct operation requires spurious signals to be >10dB below the wanted signal. |
Acquisition Time |
Typically less than 100ms to locate and lock to the carrier signal. Settling time for the demodulation and AF circuits is additional and is typically 1 second for a reading in the two thirds of the meter scale. |
Lock and Level Indicator |
A LED indicates that the instrument is locked to an input signal and that the level is within range. |
Input Impedance |
50 ohms nominal. |
Input Level |
The unit will lock and measure modulation for input carrier levels in the range 2mV to 1V. Full specification for noise, accuracy etc only applies over the input range 10mV to 1V. |
Maximum Safe Input Level |
0.5W continuous. |
Local Oscillator Feed-out |
Typically -60dBm. |
F.M. Measurement (Over Full Frequency Range) |
F.M. Ranges |
Five ranges with full scale deviations of 1kHz, 3kHz, 10kHz, 30kHz and 100kHz. |
Accuracy (+5°to +45°C) |
±2% of full scale, ±1% of reading with 1kHz tone. See Audio filter specification for additional error due to AF response. Residual FM is additional. |
Residual F.M. |
<20Hz at 100MHz
<100Hz at 500MHz
<200Hz at 1000MHz |
Distortion |
<1% at 100kHz deviation with a 1kHz tone. |
F.M. Measurement Modes |
Peak +ve. Peak -ve and Mean deviation. |
A.M. Measurement (Over Full Frequency Range) |
A.M. Ranges |
Five ranges with full scale indications of 1%, 3%, 10%, 30% and 100%. |
Accuracy <+5°to +45°C) |
±2% of full scale, ±2% of reading with a 1kHz tone. See Audio Filter specification for additional error due to audio response. Residual AM is additional. |
Residual A.M. |
Less than 0.5% with a 15kHz bandwidth selected. |
Distortion |
<1% for 80% A.M. with a 1kHz tone. |
A.M. Measurement Modes |
Peak, Trough and Mean of peak and trough. |
Audio Filters |
60kHz Bandpass Filter |
250Hz - 60kHz ±0.5dB, typically 12Hz - 72kHz ±3dB with HF roll off at 80dB/decade. |
15kHz Bandpass Filter |
250Hz - 15kHz ±0.5dB, typically 12Hz -19.5kHz ±3dB with HF roll off at 60dB/decade. |
3.5kHz Bandpass Filter |
250Hz - 3.5kHz ±0.5dB, typically 12Hz - 4.0kHz ±3dB with HF roll off at 100dB/decade. |
Psophometric Filter |
Complies with CCITT Volume V P53. |
De-emphasis network |
750µs de-emphasis, 3dB bandwidth typically 12Hz to 212Hz. HF roll off 20dB/decade. |
AF Output |
Demodulated A.F. Output |
Front Panel BNC |
A.F. Output Level |
'0'dBm corresponding to full scale deflection. |
A.F. Output Impedance |
600 ohms. |
Power Requirements |
Mains |
240V or 120V ±15%, internally selectable. |
Frequency |
48 to 60Hz. |
Power |
Approx 6VA. |
Fuse |
Rear panel 100mA quick blow fuse. |
Mechanical |
Size |
H 105mm, W 215mm, D 305mm. |
Weight |
Approx 1.7Kg
Approx 2.6Kg with battery option. |
Environmental |
Operating Temperature Range |
0° to 55°C. Full specification over the range to 45°C. |
Storage Temperature Range |
-20°C to 55°C. |
Humidity |
Max 95% RH at 30°C. |
GPIB Interface (Optional) |
Remote control and interrogation of all functions via a GPIB interface. This interface complies to the ANSI/IEEE 488-1978 standard with message formats complient with ANSI/IEEE 728-1982 standard. |
GPIB address |
0 to 15 selectable by rear panel DIL switch. |
Subsets Implemented |
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C0. |
Battery Facility (Optional) |
Discharge Time |
>8hrs, typically 10hrs for a fully charged battery. |
Recharge Time |
14hrs. |
Battery Test |
Pressing the 'Bat Chk’ switch displays the battery condition on the meter. A reading between 8 and 10 is required for operation. |
Battery Protection |
Rear panel 1A slow blow fuse. |