DC Voltage | Ranges | 400 mV, 4 V, 40 V, 400 V, 1000 V |
Accuracy all ranges: | ± (0.5 % rdg + 1 dgt) |
Resolution | 100 μV |
AC Voltage (45 Hz to 2 kHz) True RMS | Ranges | 400 mV, 4 V, 40 V, 400 V, 750 V |
Accuracy 400 mV range, 45 to 100 Hz: | ± (1.2 % rdg + 8 dgts) |
Accuracy 750 V range: 45 to 1 kHz: | ± (2.0 % rdg + 8 dgts) |
Accuracy All other ranges, 45 to 500 Hz: | ± (1.2 % rdg + 8 dgts) |
Accuracy All other ranges, 500 to 2 kHz: | ± (2.0 % rdg + 8 dgts) |
Resolution | 100 μV |
AC coupled | True RMS specified from 5 % to 100 % of range |
DC Current | Ranges | 400 μA, 4 mA, 40 mA, 300 mA, 10 A |
Accuracy 400 μA to 300 mA ranges: | ± (1.0 % rdg + 1 dgt) |
Accuracy 10 A range: | ± (2.0 % rdg + 3 dgts) |
Resolution | 0.1 μA |
AC Current (45 Hz to 1 kHz) True RMS | Ranges | 400 μA, 4 mA, 40 mA, 300 mA, 10 A |
Accuracy 400 μA to 300 mA ranges: | ± (1.5 % rdg + 8 dgts) |
Accuracy 10 A range: | ± (2.5 % rdg + 10 dgts) |
Resolution | 0.1 μA |
Resistance | Ranges | 400 Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, 40 MΩ, |
Accuracy 400, 4 k, 40 k, 400 k, 4 MΩ ranges: | ± (1.0 % rdg + 4 dgts) |
Accuracy 40 MΩ range: | ± (2.0 % rdg + 5 dgts) |
Resolution | 100 mΩ |
Capacitance | Ranges | 4 μF, 40 μF, 400 μF, 4000 μF |
Accuracy 4 μF range: | ± (5.0 % rdg + 10 dgts) |
Accuracy 40 μF, 400 uF ranges: | ± (5.0 % rdg + 5 dgts) |
Accuracy 4000 μF range: | ± (5.0 % rdg + 15 dgts) |
Resolution | 1 nF |
Temperature | Ranges | -20 ºC to 1000 ºC, -4 ºF to 1832 ºF |
Accuracy -20 ºC to 10 ºC: | ± (2.0 % rdg + 4 ºC ) |
Accuracy 10 ºC to 200 ºC: | ± (1.0 % rdg + 3 ºC ) |
Accuracy 200 ºC to 1000 ºC: | ± (3.0 % rdg + 2 ºC ) |
Accuracy -4 ºF to 50 ºF: | ± (2.0 % rdg + 8 ºF) |
Accuracy 50 ºF to 400 ºF: | ± (1.0 % rdg + 6 ºF) |
Accuracy 400 ºF to 1832 ºF: | ± (3.0 % rdg + 4 ºF) |
Resolution | 1 ºC, 1 ºF |
Frequency | Ranges 4 kHz, 40 kHz, 400 kHz, 4 MHz, 40MHz Accuracy all ranges: ± (0.1 % rdg + 3 dgts) Resolution 1 Hz | 4 kHz, 40 kHz, 400 kHz, 4 MHz, 40MHz |
Duty Cycle | Ranges | 0 to 90 % |
Accuracy 5 V logic: | ± (2.0 % rdg + 5 dgts) |
Frequency Range | 40 Hz to 20 kHz |
Continuity | Audible Indication | < 35 Ω |
Diode test | Test Current | 1.2 mA (approximate) |
Open Circuit Volts | 3.0 V dc typical |
Overload Protection | AC/DC Voltage All ranges | 1000 V dc or 750 V ac rms |
AC/DC Current μA/mA input: | 0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm) |
AC/DC Current 10 A input: | 10 A /1000 V fast-blow ceramic 10 x 38 mm (10 A for 4 minutes maximum followed by a 12 minute cooling period) |