TSX and TSX-P Technical Specifications |
OUTPUT SPECIFICATIONS |
Operating Modes: |
Constant voltage or constant current with automatic crossover |
Voltage range: |
0V to 35V (TSX3510/TSX3510P)
0V to 18V (TSX1820/TSX1820P) |
Current range: |
0A to 10A (TSX3510/TSX3510P)
0A to 20A (TSX1820/TSX1820P) |
Overvoltage protection: |
1V to 40V (35V/10A); 1V to 25V (18V/20A). |
Load regulation: |
For any load change, measured at the O/P terminals, using remote sense:
- Constant voltage: <0.01% ± 5mV
- Constant current: < 0.01% ±0.5mA |
Line regulation: |
Change in output for a 10% line change:
- Constant voltage: < 0.01% ± 5mV
- Constant current: < 0.01% ± 250uA |
Ripple & noise:
Typically (20MHz bandwidth) |
< 1mV rms; < 3mVrms max in CV mode
Typically <3mA rms; <5mArms max |
in CI mode |
HF common mode noise: |
Typically < 3mV rms, < 15mV pk. |
Transient response: |
100μs to within 50mV of setting for a 5% to 95% load change. |
Temperature coefficient:
Overvoltage protection |
Typically < 100ppm/°C. |
delay: |
<200μs |
Protection functions: |
Overvoltage trip.
Regulator over temperature.
Sense miswiring |
Status indication: |
LED indicators for Output ON, constant voltage mode, constant current mode, current meter damping, remote operation, and LAN status; trip & error messages on display |
Output terminals: |
Front panel: Universal 4mm safety binding posts on 19mm (0·75") spacing Rear panel: Terminal block screw terminals. |
Output protection: |
Forward protection by Over-Voltage-Protection (OVP) trip. Output will withstand an applied forward voltage of up to 50V (35V/10V), 35V (18V/20A) Reverse protection by diode clamp for reverse currents up to 3A. |
METER SPECIFICATIONS |
Meter types: |
Dual 4 digit meters with 12·5mm (0.5")LEDs. Reading rate 4Hz |
Meter resolutions: |
10mV,10mA. |
Meter accuracies: |
Voltage 0.2% of reading ± 1 digit,
Current 0.5% of reading ± 1 digit. |
GENERAL |
Input voltage range: |
110V – 240V AC ± 10%, 50/60Hz.
Installation Category II. |
Power requirement: |
600VA max. |
Electrical safety: |
Complies with EN61010-1. |
EMC: |
Complies with EN61326-1 |
Temperature: |
Operating: +5°C to +40°C, 20% to 80%RH. Storage: -40°C to +70°C. |
Cooling: |
Fan-less convection cooling. |
Size: |
200 x 140 x 385mm (WxHxD); Half rack width x 3U height. |
Weight: |
4.5kg (TSX versions) 5.0kg (TSX-P versions) |
Rack mount Options: |
3U Rack Mount for half rack width instruments, (RM300A) |
FRONT PANEL CONTROLS - TSX |
Voltage setting: |
Via single rotary controls for coarse and fine control. |
Current setting: |
Via single turn semi-logarithmic rotary control. |
Overvoltage setting: |
Via screwdriver adjustable preset potentiometer |
Output On/Off: |
Latching push-push switch operating electronic power control. |
FRONT PANEL CONTROLS - TSX-P |
Voltage setting: |
Direct keyboard entry or quasi-analog rotary control. |
Current setting: |
Direct keyboard entry or quasi-analog rotary control. |
Overvoltage setting: |
Direct keyboard entry |
Output On/Off: |
Push button control with indicator lamp operating electronic power control. |
Delta Mode: |
Increase or decrease voltage or current in user- selectable steps. |
Store/Recall: |
Store and recall voltage, current and OVP levels from non-volatile memory (25 memories). |
Address Setting: |
Set GPIB address |
Damping ON/OFF: |
Sets current meter damping ON or OFF |
NOTE: all voltage and current levels set via the keyboard are displayed on a separate 0.3” 4 digit display. This entry preview system ensures that the user can observe the value entered before it is effected thus avoiding possible error. The display is also used for setting additional functions and for displaying watts. |
DIGITAL INTERFACES - TSX-P |
Operational Functions: |
Set voltage, set current, set OVP; set output on/off; read output voltage; read output current; read output power. |
RS232: |
9600 baud, 9 pin D-connector (female). |
GPIB (Optional): |
Conforming with IEEE488.1 & IEEE488.2. |
USB Interface: |
Standard USB 2.0 hardware connection. Implemented as a Virtual COM Port. |
LAN Interface: |
LAN Ethernet 100/10base-T hardware connection. 1.4 LXI Core 2011 |
Setting resolution: |
Voltage - 10mV. Current - 10mA. |
Setting accuracy: |
Voltage - ±(0.1% + 10mV) Current - ±(0.2% + 20mA) |
Readback resolution: |
Voltage - 10mV. Current - 10mA. |
Readback accuracy: |
Voltage - ±(0.2% + 1 digit) Current - ±(0.5% +1 digit |
Operating software: |
Software for operating the PSUs under GPIB or RS232 control is available including a Labwindows* driver. |
RESPONSE TIMES: |
Remote Command
Processing Time: |
Typically <50ms between receiving the command terminator for a step voltage change at the instrument and the output voltage beginning to change. |
Power Supply: |
An internal time constant, T, (typically 22ms) governs the settling time of a step voltage increase. |
NOTE: Settling time to within 1% of the step change = 4.6T, to 0.1% = 6.9T, to 0.01% = 9.2T; for example, after a 10V step the output will be within 1 digit (10mV = 0.1%) of its new value in typically 150ms. For load current of 1 Amp or more, settling times for downward steps will be very similar; however, response times will be longer at low loads |