OUTPUT SPECIFICATIONS (each output) |
Voltage/Current/Power Levels |
Voltage Range |
0V to 60V |
Current Range |
0A to 20A |
Power Range |
Up to 420W subject to power envelope |
Output Setting & Control |
Voltage Setting |
By coarse and fine controls |
Current Setting |
By single logarithmic control |
Output Mode |
Constant voltage or constant current with automatic cross-over CC indicator lit in constant current mode |
Output Switch |
Electronic, non isolating. Preset voltage and current limit displayed when Output is off. Output rise time no load <10ms |
View Settings |
With the output On, the meters show actual voltage and current. The preset levels can be viewed and adjusted at any time by pressing the View Settings button |
Status Indication |
LED indication of Output On, V/I Limits, CV, CI, Power Limit, Remote, LAN status. Message on meter display for trip condition |
S-Lock |
(Settings Lock) Voltage and current settings can be locked by a single button press. Lock accuracy is equal to the meter accuracy |
Output Performance |
Ripple & Noise |
Typically <3mV rms, <15mV pk-pk, (5mV rms max.) in CV mode |
Load Regulation |
Voltage - <0·01% of maximum output for any load change within the PowerFlex envelope (remote sense connected)
Current - <0·05% of maximum output for any load change within the PowerFlex envelope |
Line Regulation |
Voltage - <0·01% of maximum output for a 10% line change
Current - <0·01% of maximum output for a 10% line change |
Transient Response |
<250µs to within 50mV of setting for a 5% to 95% load change |
Temp. Coefficient |
Typically <100ppm/°C |
Output Protection |
Output Protection |
Forward protection by Over-voltage Protection (OVP) trip
Reverse protection by diode clamp for currents to 3A |
OVP Setting/Range |
Via screwdriver adjustable preset on front panel. Range 1V to 66V |
Over-temperature |
Output trips off for over-temperature. |
Safety Interlocks |
Operations that could cause an unexpected change in voltage or current settings are interlocked with the output switch |
METER SPECIFICATIONS (each output) |
Display Type |
Dual 4-digit meters, 10mm (0·39”) LED. |
Voltage Meter |
Resolution/ Accuracy |
10mV / ± 0·1% of reading ± 2 digits |
Current Meter |
Resolution/ Accuracy |
10mA / ± 0·3% of reading ± 20mA |
GENERAL SPECIFICATIONS |
Input |
AC Input |
110 to 240 volts ±10% 50/60Hz. Installation Category II |
Input Power |
Single - 625VA, Dual - 1250VA max. |
Temperature & Environmental |
Operating Range |
+5ºC to +40ºC, 20% to 80% RH |
Storage Range |
-40ºC to + 70ºC |
Environmental |
Indoor use at altitudes up to 2000m, Pollution Degree 2 |
Cooling |
Rear discharge variable speed fan. |
Safety & EMC |
Safety |
Complies with EN61010-1 |
EMC |
Complies with EN61326 |
Physical Size |
210 x 130 mm (½ rack 3U) x 377mm
(size excludes feet, knobs and terminals). |
Weight |
6.3kg (13.9lb) |