Earth continuity / resistance measurement |
Display range |
0.00 to 199Ω |
Measurement range |
0.01 to 199Ω |
Accuracy |
±(2% rdg + 5d) |
Resolution |
0.01O maximum |
Open circuit test voltage |
4VDC, nominal |
Test leads zero |
Zero up to 10Ω, by Zero button |
Number of measurements |
5,000 x 1 second tests |
Audible / visible warning |
= 30VAC/DC at inputs |
User protection |
Test inhibited if = 30VAC/DC at inputs |
Insulation resistance (auto short circuit test) |
Display range |
0.05 - 200MΩ |
Measurement range |
0.05 - 200MΩ |
Accuracy |
±(5% rdg + 5d)
0.05 - 100MΩ
±(10% rdg + 5d)
101 - 200MΩ |
Resolution |
0.01MΩ maximum |
Open circuit test voltage |
250, 500, 1000V
(as per IEC 61557-2) |
Test current |
1mA nominal as per IEC 61557-2 |
Short circuit test current |
<2mA |
Number of measurements |
5,000 x 1 second tests |
Audible / visible warning |
= 30VAC/DC at inputs |
User protection |
Test inhibited if = 30VAC/DC at inputs |
Insulation resistance (point to point) |
Display range |
0.05 to 300MΩ |
Measurement range |
0.05 to 300MΩ |
Accuracy |
±(5% rdg + 5d) |
Resolution |
0.01MΩ maximum |
Open circuit test voltage |
250, 500, 1000V
(as per IEC 61557-2) |
Short circuit test current |
<1mA |
Number of measurements |
5,000 x 1 second tests |
Audible / visible warning |
= 30VAC/DC at inputs |
Circuitry protection |
Test inhibited if = 30VAC/DC at inputs |
Voltage measurement (via 4mm probes) |
Display range |
30V – 440VAC/DC |
Measurement range |
30V – 440VAC/DC |
Resolution |
1V |
Accuracy |
±(5% rdg + 2d) |
Vo/c voltage measurement (via PV test leads) |
Display range |
0.0V – 1000VDC |
Measurement range |
5.0V – 1000VDC |
Resolution |
0.1V |
Accuracy |
±(0.5% rdg + 2d) |
Enunciators |
DC voltage polarity correct or
reversed |
Is/c current measurement (via PV test leads) |
Display range |
0.0A – 15.0ADC |
Measurement range |
0.5A – 15.0ADC |
Resolution |
0.1A |
Accuracy |
±(1% rdg + 2d) |
Operating current (via DC current clamp) |
Display range |
0.0A – 40.0A AC/DC |
Measurement range |
0.1A – 40.0A AC/DC |
Resolution |
0.1A |
Accuracy |
±(5% rdg + 2d) |
DC power |
Display range |
0.0W – 40.0kW |
Measurement range |
10W – 40.0kW |
Resolution |
10W max |
Accuracy |
±(6% rdg + 2d) |
I-V curve |
Maximum power dissipation |
10kW |
Number of points |
Dynamic up to 128 |
MPP calculation max error |
±(1.5% rdg + 40w) |
Case dimensions and weight |
Weight |
2.3lb (unit) |
Dimensions |
10.4 x 4.2 x 2.3” |
Display |
Custom LCD with backlight |
Power source |
6 x 1.5V AA cells |
Battery life |
>1000 test sequences |
Auto power down |
User programmable |
Onboard memory |
Up to 999 complete test datasets |
Connectivity |
USB download to PC (CSV format) |
Wireless ‘Solarlink TM ’ to Survey 200R (915MHz) (range c. 30m / 100ft) |
NFC transfer of data to PVMobile Android app |
iOS devices not supported |
App compatibility |
Compatible with Android version 4.2 (Jelly Bean) or later |
iOS devices not supported |
Software compatibility |
Compatible with SolarCert Elements v2 software or later |
(English language only) |