Signal frequency |
|
1 GHz |
Measurand |
|
RF linearity relative power measurement (ratio of two consecutive power measurements) |
Test port power range |
1 GHz |
-60 dBm to +27 dBm |
|
range 1, internal RF attenuator on |
-30 dBm to +27 dBm |
|
range 2, internal RF attenuator of |
-60 dBm to -3 dBm |
|
DC |
-10 V to +10 V |
Test port connector |
|
N female |
Output impedance |
|
50 Ω |
Insertion loss |
between RF input and RF output |
8.5 dB (meas.) |
Uncertainty for relative power measurements referenced to the power level of 0 dBm (from −21 dBm to +21 dBm) |
|
0 dB to 21 dB level difference |
0.005 dB |
Uncertainty for relative power measurements between any two power levels within one power range (from −30 dBm to +27 dBm or −60 dBm to −3 dBm) |
|
level difference A |
0.0002 × A + 0.004 dB |
|
0 dB to 5 dB |
0.005 dB |
|
> 5 dB to 10 dB |
0.006 dB |
|
> 10 dB to 15 dB |
0.007 dB |
|
> 15 dB to 20 dB |
0.008 dB |
|
> 20 dB to 30 dB |
0.010 dB |
|
> 30 dB to 40 dB |
0.012 dB |
|
> 40 dB to 50 dB |
0.014 dB |
|
> 50 dB to 57 dB |
0.016 dB |
Uncertainty for relative power measurements between any two power levels with range switching as well as calibration of the adjacent range at a level of −30 dBm (test port power range from −60 dBm to +27 dBm) |
|
level difference A |
0.0002 × A + 0.006 dB |
|
0 dB to 5 dB |
0.007 dB |
|
> 5 dB to 10 dB |
0.008 dB |
|
> 10 dB to 15 dB |
0.009 dB |
|
> 15 dB to 20 dB |
0.010 dB |
|
> 20 dB to 30 dB |
0.012 dB |
|
> 30 dB to 40 dB |
0.014 dB |
|
> 40 dB to 50 dB |
0.016 dB |
|
> 50 dB to 60 dB |
0.018 dB |
|
> 60 dB to 70 dB |
0.020 dB |
|
> 70 dB to 80 dB |
0.022 dB |
|
> 80 dB to 87 dB |
0.024 dB |
Uncertainty for relative power measurements at DC depends on the connected test and measurement instruments |
Resolution bandwidth (RBW) |
single sideband (SSB) mode |
10 Hz to 40 MHz |
|
recommended for linearity measurement |
1 kHz |
Displayed average noise level (DANL) |
test port power range 1, 1 GHz |
< 133 dBm (1 Hz) |
|
test port power range 2, 1 GHz |
< 103 dBm (1 Hz) |
LO phase noise at 1 kHz offset, measurement bandwidth 1 Hz, measured at LO I/O connector |
|
1 GHz |
–92 dBc (nom.) |
LO leakage at test port (LO frequency and frequency of harmonics) |
|
test port power range 1, 1 GHz |
-84 dBm (nom.) |
|
test port power range 2, 1 GHz |
-86 dBm (nom.) |
Other characteristics |
Sensor type |
|
R&S®NRQ6 frequency selective power sensor |
Measurement function |
recommended |
continuous average |
Continuous average function |
measurand |
average power over acquisition interval |
|
aperture |
8.3 ns to 30 s (depending on RBW) |
|
duty cycle correction |
0.001% to 100.0% |
|
capacity of measurement buffer |
1 reading to 8192 readings |
Triggering |
supported measurement functions |
continuous average |
|
source |
INTernal: internal test signal
EXTernal[1]: host interface trigger signal (8-pin male M12 connector)
EXTernal2: coaxial trigger I/O (SMA (f) jack)
BUS: remote control event (*trg) |
Averaging filter |
parameters |
|
|
supported measurement functions |
continuous average, trace |
|
averaging count |
1 to 65536 |
|
result output |
|
|
moving mode |
continuous result output, independent of averaging count |
|
repeat mode |
final result only |
Internal reference frequency |
accuracy |
±1 × 10–6 |
Intermediate frequency (IF) |
RBW ≤ 40 MHz |
20 MHz to 30 MHz |
Host interface (8-pin male M12 connector)
USB interface to PC via R&S®NRP-ZKU interface cable (requires additional PoE+ power supply at LAN interface)
USB interface to PC via R&S®NRP-ZK6 interface cable + R&S®NRP-Z5 USB sensor hub (requires additional PoE+ power supply at LAN interface)
Interface to R&S®NRX power meter via R&S®NRP-ZK6 or R&S®NRP-ZK8 interface cable |
|
mechanical |
8-pin male M12 connector (A-coded) |
|
power supply |
+5 V/0.1 A (USB low-power device; requires additional PoE+ power supply) |
|
speed |
high-speed and full-speed mode in line with USB specification |
|
remote control protocols |
USB test and measurement device class (USBTMC) |
|
trigger input EXTernal[1] |
differential (0 V/+3.3 V) |
|
R&S®NRX common time base clock (only available with R&S®NRP-ZK8 interface cable) |
|
|
signal level |
LVDS |
|
input frequency |
20 MHz |
|
permissible total cable length |
≤ 5 m |
Ethernet interface (LAN PoE+) |
mechanical |
RJ-45 jack |
|
power supply |
power over Ethernet (PoE+) class 4 |
|
speed |
10/100/1000 Mbit/s |
|
remote control protocols |
VXI-11, high-speed LAN instrument protocol (HiSLIP), SCPI-RAW (port 5025) |
|
permissible cable length |
≤ 100 m |
Trigger 2 I/O (TRIG2) |
mechanical |
SMA (f) jack |
|
impedance |
|
|
input |
10 kΩ or 50 Ω (software controlled) |
|
output |
50 Ω |
|
signal level |
|
|
input |
compatible with 3 V or 5 V logic, max. –1 V to +6 V |
|
output |
≥ 2 V into 50 Ω load, max. 5.3 V |
Reference clock I/O (REF) |
mechanical |
SMA (f) jack |
|
impedance |
|
|
input/output |
50 Ω |
|
signal level |
|
|
input |
≥ –10 dBm |
|
output |
≥ +7 dBm |
|
frequency |
|
|
input |
10 MHz |
|
output |
10 MHz |
Sample clock I/O (CLK) |
mechanical |
SMA (f) jack |
|
impedance |
|
|
input/output |
50 Ω |
|
signal level |
|
|
output |
≥ -10 dBm |
|
frequency |
|
|
output |
119 MHz to 121 MHz |
Local oscillator I/O (LO) |
mechanical |
SMA (f) jack |
|
impedance |
|
|
input/output |
50 Ω |
|
signal level |
|
|
input |
≥ -5 dBm |
|
output |
≥ 0 dBm |
|
frequency |
|
|
input/output |
70 MHz to 6.03 GHz |
General data |
Temperature |
operating temperature range |
-19.7°C to +26.3°C |
|
permissible temperature range |
0°C to +40°C |
|
storage temperature range |
-40°C to +70°C |
Climatic resistance |
damp heat |
+25°C/+40°C cyclic at 95% relative humidity with restrictions: noncondensing, in line with EN60068-2-30 |
Mechanical resistance |
vibration |
|
|
sinusoidal |
5 Hz to 55 Hz, 0.15 mm amplitude, 1.8 g at 55 Hz, 55 Hz to 150 Hz, 0.5 g constant, in line with EN60068-2-6 |
|
random |
8 Hz to 500 Hz, 1.2 g (RMS), in line with EN60068-2-64 |
|
shock |
45 Hz to 2 kHz, max. 40 g shock spectrum, in line with MIL-STD-810E, method 516.4, procedure I |
Air pressure |
operating |
795 hPa (2000 m) to 1060 hPa |
|
transport |
566 hPa (4500 m) to 1060 hPa |
Electromagnetic compatibility |
|
complies with harmonized standards: EN61326-1, EN61326-2-1, EN55011 (class B) |
Calibration interval |
recommended |
2 years |
Dimensions |
W × H × D |
233 mm × 103 mm × 372 mm (9.18 in × 4.06 in × 14.65 in) |
Weight |
|
4.2 kg (9.26 lb) |