Rohde and Schwarz NRP-Z81 Specifications |
Frequency Range |
R&S®NRP-Z81 |
50 MHz to 18 GHz |
Impedance Matching (SWR) |
50 MHz to 2.4 GHz |
< 1.16 (1.11) |
( ): +15 °C to +35 °C |
> 2.4 GHz to 8.0 GHz |
< 1.20 (1.18) |
> 8.0 GHz to 18.0 GHz |
< 1.25 (1.23) |
Power Measurement Range |
Continuous average |
1 nW to 100 mW (-60 dBm to +20 dBm) |
Burst |
Full video bandwidth |
20 µW to 100 mW (-17 dBm to +20 dBm) |
300 kHz |
4 µW to 100 mW (-24 dBm to +20 dBm) |
Trace, timeslot/gate |
20 nW to 100 mW (-47 dBm to +20 dBm) |
Statistics |
4 µW 25 to 100 mW (-24 dBm to +20 dBm) |
Max. Power |
Average power |
0.2 W (+23 dBm), continuous |
Peak envelope power |
1.0 W (+30 dBm) for max. 1 µs |
Dynamic Response |
Video bandwidth |
≥ 30 MHz 26 |
Single-shot bandwidth |
≥ 30 MHz 26 |
Video bandwidth setting |
Full (≥ 30 MHz), 5 MHz, 1.5 MHz, 300 kHz |
Rise time 10 %/90 % |
Full video bandwidth |
≤ 13 ns 26 (f ≥ 500 MHz) |
< 40 ns 26 (f < 500 MHz) |
5 MHz |
< 75 ns |
1.5 MHz |
< 250 ns |
300 kHz |
< 1.2 µs |
Detectable burst width |
≥ 50 ns 26 (f ≥ 500 MHz, full video bandwidth) |
Overshoot |
≤ 5 % |
Acquisition |
Sample rate [period] |
Full video bandwidth |
80 × 10 6 s -1 [12.5 ns] |
5 MHz |
40 × 10 6 s -1 [25.0 ns] |
1.5 MHz |
10 × 10 6 s -1 [100 ns] |
300 kHz |
2.5 × 10 6 s -1 [400 ns] |
Capture length |
50 ns to 1 s (depending on meas. function) |
Time base accuracy |
±50 ppm |
Time base jitter |
< 1 ns |
Triggering |
Internal |
Threshold level range |
-30 dBm to +20 dBm
(usable from -22 dBm with full video bandwidth) |
Threshold level accuracy |
Identical to uncertainty for absolute power measurements |
Threshold level hysteresis |
0 dB to 10 dB |
Dropout 5 |
0 s to 10 s |
External |
see R&S®NRX base unit, R&S®NRP-Z3 USB adapter cable or R&S®NRP-Z5 USB sensor hub |
Slope (external, internal) |
pos./neg. |
Delay |
-51.2 µs to +10 s |
Hold-off |
0 s to 10 s |
Resolution (delay, hold-off, dropout) |
Sample period |
Source |
Internal, external, immediate, bus, hold |
5 Time span prior to triggering, where the trigger signal must be entirely below the threshold level in the case of a positive slope and vice versa in the case of a negative slope. |
25 With full video bandwidth. Reduce the specified minimum levels according to the reduction of sampling noise at lower bandwidths. |
26 Specifications are valid from +15 °C to +50 °C ambient temperature. Below +15 °C, video bandwidth and single-shot bandwidth continuously decrease down to 20 MHz (typical) at 0 °C. Accordingly, the sensor rise time increases up to 50 ns for signals below 500 MHz and up to 20 ns for higher frequencies (typical at 0 °C). |