Memmert WTB 11 Specifications |
Temperature |
Temperature sensor |
digital temperature sensor, mounted on the outside of the bath |
Temperature |
Setting Celsius / Fahrenheit |
Working temperature range |
at least 5 above ambient to +100 (depending on ambient pressure) |
Setting temperature range |
+10 to +100 °C |
Setting accuracy temperature |
0.1 °C |
Control technology |
Calibration |
2-point calibration |
Controller |
intuitive 3.5" colour graphic display with touch screen |
Controller |
digital display of set and actual temperature and of (remaining) programme time |
Safety |
Temperature control |
two-stage safety system to prevent overtemperature |
Temperature control |
individually adjustable overtemperature alarm, visual and acoustic |
Temperature control |
Independent temperature limiter to switch off the heating in case of failure |
Autodiagnostic system |
microprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator |
Heating concept |
Heating |
corrosion-proof large-area heating on the outside of the interior |
Standard equipment |
Drain system |
innovative drain system - easy to empty |
Timer |
digital timer from 1 min. up to 99 hours 59 min., visual and acoustic |
Stainless steel interior |
Interior |
Deep-drawn stainless steel tray (stainless steel W.St. 1-4301), without corners and edges for easy cleaning incl. central drain (no disturbing installations) |
Dimensions W x H x D in mm |
w(A) x h(C) x d(B): 267 x 200 x 207 mm |
Volume |
10 l |
Liquid level min. |
40 mm |
Textured stainless steel casing |
Dimensions |
w(D) x h(E2) x d(F): 349 x 398 x 289 mm |
Electrical data |
Voltage |
230 V, 50/60 Hz |
Electrical load |
approx. 1000 W |
Voltage |
115 V, 50/60 Hz |
Electrical load |
approx. 1000 W |
Ambient conditions |
Installation |
The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm. |
Ambient temperature |
+5 °C to +40 °C |
Humidity rh |
max. 80 %, non-condensing |
Overvoltage category |
II |
Pollution degree |
2 |
Packing/shipping data |
Transport information |
The appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
w x h x d: 570 x 450 x 380 mm |
Net weight |
approx. 7 kg |
Gross weight carton |
approx. 12 kg |
Standard units are safety-approved and bear the test marks |
CE, EAC |