Measurement |
Range |
0.1 µΩ to 999.9 mΩ
(Subject to supply voltage and leads used) |
Accuracy |
Voltage |
±0.5% ± 0.1 mV |
Current |
±0.5% ± 0.1 A |
Resistance |
Better than 1% from
300 µΩ to 100 mΩ
Better than ±2% from
100 µΩ to 300 µΩ |
Current Lead Resistance (Megger supplied leads) |
2 x 5 m 25 mm2 current leads |
8 m |
2 x 5 m 50 mm2 current leads |
4 mΩ |
2 x 10 m 70 mm2 current leads |
5.4 mΩ |
2 x 15 m 95 mm2 current leads |
6 mΩ |
Maximum Continuous Test Time |
|
More than 60 seconds at 200 A @ 20°C ambient. |
Power Supply for DLRO200 for full output (unsmoothed d.c.) |
|
100 to 265 V 50/60 Hz with a load
less than 19 mΩ (supply >207 V rms), or
11 mΩ (115 V rms) including current leads |
Power Supply for DLRO200-115 V for full output (additional smoothing) |
|
100 to 130 V 50/60 Hz with a load
less than 11 mΩ including current leads |
Test Modes |
Manual, Auto, Continuous. |
Test Time |
10 seconds NORMAL /AUTO mode.
Refreshed every 2 seconds in CONTINUOUS mode |
Display |
Large, high resolution backlit liquid crystal display |
Warnings Current flowing |
- LED. Other warnings are
shown on the lcd display |
Data Transfer Real time or batch download via RS232 using Download Manager. |
Storage Capacity |
|
300 result sets and memo, battery backed for 10 years. |
Memo field |
160 characters max. |
Test Current |
Range |
10 A to 200 A d.c. * |
Accuracy |
±2% ±2 A |
Voltmeter input |
impedance |
>200 kΩ |
Hum rejection |
5 V rms 50 Hz/60 Hz |
Temperature |
Operation |
-10 to +50°C (-14 to +122°F) |
Storage |
-25 to +65°C (-13 to 149°F) |
Calibration |
20°C |
Co-efficient |
<0.05% per °C |
Max. Humidity |
95% RH non-condensing |
Maxi Altitude |
2000 m |
Safety |
IEC61010-1 |
EMC |
IEC61326-1 |
Dimensions |
410 x 250 x 270 mm
(16.4 x 9.84 x 10.63 ins) |
Weight |
14.5 kg (excluding test leads)
(31.97 lbs) |