Temperature Sensor |
HiTemp140X2- FPST: Two Flexible RTD Probes
HiTemp140X2- FPST-PT: Flexible RTD Probe and Bendable RTD Probe
HiTemp140X2-TD-PT: Rigid RTD Probe with a Bendable RTD Probe
HiTemp140X2-TD-FPST: Rigid RTD Probe with a Flexible RTD Probe |
Probe Measurement |
Range Rigid Probe: -200 °C to +260 °C (-328 °F to +500 °F)
Bendable Probe: -200 °C to +350 °C (-328 °F to +662 °F)
Flexible Probe: -60 °C to +260 °C (-76 °F to +500 °F)
*BODY OF LOGGER CANNOT EXCEED 140 °C |
Temperature Resolution |
0.01 °C (0.02 °F) |
Calibrated Accuracy |
±0.1 °C (+20 °C to +140 °C); 0.18 °F (+68 °F to +284 °F)
±0.3 °C (-20 °C to 19.99 °C); 0.54 °F (-4 °F to +67.98 °F)
±0.4 °C (-40 °C to -20.01 °C); 0.72 °F (-40 °F to +68.018 °F) |
Start Modes |
Software programmable immediate start
Delay start up to 18 months in advance |
Stop Modes |
Manual or Timed (specific date and time) |
Trigger Settings |
High and Low limits may be set. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers. (Triggering on channel #1 only) |
Readings in Trigger Settings Mode |
16,383 readings |
Real Time Recording |
May be used with PC to monitor and record data in real time |
Password Protection |
An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password |
Memory |
32,767 readings |
Wrap Around |
Yes |
Reading Rate |
1 reading every second up to 1 reading every 24 hours |
Calibration |
Digital calibration through software |
Calibration Date |
Automatically recorded within device |
Battery Type |
3.6 V high-temperature lithium battery included; user replaceable |
Battery Life |
1 year typical (1 minute reading rate at 25 °C/ 77 °F) |
Data Format |
Date and time stamped °C, K, °F or °R |
Time Accuracy |
±1 minute/month at 25 °C (77 °F) Extended Operation: ±20 minutes/month at 140 °C (±450 ppm) |
Computer Interface |
IFC400 OR IFC406 USB docking station required; 125,000 baud |
Operating System Compatibility |
Windows XP SP3 or later |
Software Compatibility |
MadgeTech Standard Software version 4.2.1.0 or later MadgeTech Secure Software version 4.2.0.0 or later |
Operating Environment |
-40 °C to +140 °C (-40 °F to +284 °F) 0 %RH to 100 %RH, 0.002 PSIA to 100 PSIA |
IP Rating |
IP68 |
Dimensions (Body) |
1.89 in x 0.97 in x 0.97 in (48 mm x 24.6 mm x 24.6 mm) |
Dimensions (Probe) |
6 inch Flexible Probe: 6 in x 0.13 in (152 mm x 3.2 mm)
12 inch Flexible Probe: 12 in x 0.13 in (305 mm x 3.2 mm)
36 inch Flexible Probe: 36 in x 0.13 in (914.4 mm x 3.2 mm)
72 inch Flexible Probe: 72 in x 0.13 in (1829 mm x 3.2 mm)
1 Inch PT Bendable Probe: Probe tip: 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.)
Bendable portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
5 Inch PT Bendable Probe:
Probe tip: 4.8 in x 0.125 in dia. with 1 in x 0.188 in dia. handle (121 mm x 3.2 mm dia. with 25 mm x 4.8 mm dia. handle)
Bendable portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
TD Rigid Probe: 2.0 in x 0.125 in dia. (0.188 in transitional dia.) 51 mm x 3.2 mm dia. (4.8 mm transitional dia.) |
Weight |
HiTemp140X2-FPST: 4.1 oz (115 g)
HiTemp140X2-FPST-PT: 3.9 oz (110 g)
HiTemp140X2-TD-PT: 3.0 oz (85 g)
HiTemp140X2-TD-FPST: 3.5 oz (100 g) |
Material |
Body: 316 Stainless Steel, PEEK
Bendable Probe: 316 Stainless Steel
Flexible Probe: PFA Insulated Cable with Stainless Steel Tip |
Approvals |
CE |