Analog - Vertical |
Bandwidth (@ 50Ω) |
200 MHz |
200 MHz |
350 MHz |
500 MHz |
750 MHz |
Rise time |
1.75 ns typical |
1.75 ns typical |
1 ns typical |
800 ps typical |
550 ps typical |
Input Channels |
2 |
4 |
Vertical Resolution |
8-bits; up to 11-bits with enhanced resolution (ERES) |
Sensitivity |
50 Ω: 1mV/div - 1 V/div; 1 MΩ: 1 mV/div - 10 V/div |
DC Gain Accuracy |
±1.0% of full scale (typical); ±1.5% of
full scale = 10 mV/div (warranted) |
BW Limit |
20 MHz |
20 MHz, 200 MHz |
Maximum Input Voltage |
50 Ω: 5 Vrms, 1 MΩ: 250 V max. (DC + Peak AC =10 kHz) |
Input Coupling |
50 Ω: DC, GND; 1 MΩ: AC, DC, GND |
Input Impedance |
50 Ω ±2.0%, 1 MΩ ±2.0% || 16 pF |
Analog - Acquisition |
Sample Rate (Single-shot) |
5 GS/s (10 GS/s Interleaved) |
Sample Rate (Repetitive) |
50 GS/s |
Record Length |
10 Mpts/Ch (all channels),
20 Mpts (interleaved) |
Acquisition Modes |
Real Time, Roll, RIS (Random Interleaved Sampling),
Sequence (Segmented Memory up to 1,000 segments with 1µs minimum intersegment time) |
Real Time Timebase Range |
2 ns/div - 50 s/div |
1 ns/div - 50 s/div |
RIS ModeTimebase Range |
2 ns/div - 10 ns/div |
1 ns/div - 10 ns/div |
Roll Mode Timebase Range |
Up to 50 s/div (roll mode is user selectable at = 100 ms/div) |
Timebase Accuracy |
±10 ppm measured over > 1ms interval |
Probes |
Standard Probes |
One PP019 (5mm) per channel |
One PP020 (5mm) per channel |
Probing System |
BNC and Teledyne LeCroy ProBus for Active voltage, current and differential probes |
Display System |
Display Size |
10.1” Wide TFT-LCD Touch-Screen |
Display Resolution |
1024 x 600 |
Connectivity |
Ethernet Port |
10/100 Base-T Ethernet interface (RJ-45 connector) |
Removable Storage |
(1) MicroSD Port - 8 GB MicroSD card installed standard) |
USB Host Ports |
(4) USB Ports Total – (2) Front USB Ports |
USB Device Port |
(1) USBTMC |
GPIB Port (Optional) |
Supports IEEE – 488.2 |
External Monitor Port |
Standard DB-15 connector (support resolution of 1024x600) |
Remote Control |
Via Windows Automation, or via Teledyne LeCroy Remote Command Set |
Network Communication Standard |
GPIB IEEE-488.2, VXI-11 and VICP, USBTMC/ USB488 |
Power Requirements |
Voltage |
100 - 240 VAC ± 10% at 50-60 Hz +/-5%; 100 - 120 VAC ± 10% at 400 Hz +/- 5%; Automatic AC Voltage Selection |
Power Consumption (Nominal) |
100 W / 100 VA |
Power Consumption (Max) |
150 W / 150 VA (with all PC peripherals, digital lead set and active probes connected to 4 channels) |
Environmental |
Temperature |
Operating: 0 ΩC to 50 ΩC; Non-Operating: -30 ΩC to 70 ΩC |
Humidity |
Operating: 5% to 90% relative humidity (non-condensing) up to = 30 ΩC, Upper limit derates to 50% relative humidity (non-condensing) at +50 ΩC
Non-Operating: 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F |
Altitude |
Operating: 3,048 m (10,000 ft) max at = 25C; Non-Operating: Up to 12,192 meters (40,000 ft) |
Physical |
Dimensions (HWD) |
10.63”H x 14.96”W x 4.92”D (270 mm x 380 mm x 125 mm) |
Weight |
4.81 kg (10.6 lbs) |
Regulatory |
CE Certification |
Low Voltage Directive 2006/95/EC; EN 61010-1:2010, EN 61010-2-030:2010
EMC Directive 2004/108/EC; EN 61326-1:2013, EN61326-2-1:2013; RoHS2 Directive 2011/65/EU |
UL and cUL Listing |
UL 61010-1, UL 61010-2-030:2010, 3rd Edition; CAN/CSA C22.2 No. 61010-1-12 |
Calibrated |
Commercial NIST Traceable Calibration with Certificate |
Certificate of Calibration |
Yes |
Warranty |
3 Years |
Function |
Math Functions Include Sum, Difference, Product, Ratio, Absolute Value, Averaging (summed and continuous), Derivative, Envelope, Enhanced Resolution (to 11-bits), Floor, Integral, Invert, Reciprocal, Rescale (change scale and units), Roof, SinX / X, Square, Square Root, Trend, Zoom and FFT (up to 1 Mpts with power spectrum and magnitude outputs and rectangular, VonHann, FlatTop, Hamming and Blackman Harris windows). 2 single operator math functions may be defined at a time. |
Included with Standard Configurations |
÷10 Passive Probe (Total of 1 Per Channel), Getting Started Guide, Microsoft Windows Embedded Compact 7,
Commercial NIST Traceable Calibration with Certificate, Power Cable for the Destination Country, 3-year Warranty |
I2C, SPI, UART and RS-232 Trigger (optional) |
I2C |
Trigger on START, ReSTART, STOP, ADDR, DATA, ADDR+DATA, Data Length, or Missing ACK. 7 or 10-bit ADDR are supported with full Read, Write, or R/W = “Don’t Care”. Choose to trigger on ADDR values that include/don’t include R/W bit in the value.DATA conditions support <=, <, =, >, >=, <>, in range, out of range, and don’t care. DATA can be setup in either Hexadecimal (0 to 12 bytes supported and can be defined by nibble) or Binary (any combination of 0, 1, or X for 1-96 bits). DATA patterns can be set to start on any bye in a 2048-byte window (EEPROM mode only). ACK condition can be selected as ACK, NO ACK, and DON’T CARE. Full range of bit rates supported for Standard, Fast, Fast-Mode Plus, and High speed mode. Bit rate is auto-detected. Trigger on any analog or digital channel. Triggering on EXT or EXT/5 is not supported. |
SPI |
Trigger on DATA.DATA can be setup in either Hexadecimal (0 to 12 bytes supported and can be defined by nibble) or Binary (any combination of 0, 1, or X for 1-96 bits). All bit rates are supported and are auto-detected. Trigger on any analog or digital channel. Triggering on EXT or EXT/5 is not supported. |
UART / RS-232 |
Trigger on DATA or Parity ERROR.DATA conditions support <=, <, =, >, >=, <>, in range, out of range, and don’t care. DATA can be setup in either Hexadecimal (0 to 12 bytes supported and can be defined by nibble) or Binary (any combination of 0, 1, or X for 1-96 bits). Data position can be specified anywhere in a 2048 byte sequence. All bit rates are supported between 300 b/s and 10 Mb/s. Trigger on any analog or digital channel. Triggering on EXT or EXT/5 is not supported. |