MIL-STD-1553 |
|
Definition |
Protocol Setup |
Select data source. |
|
Decode Capability |
Format |
Hexadecimal, Binary |
Decode Setup |
Threshold definition required. Default is to percent amplitude. |
Decode Input |
Any analog Channel, Memory or Math trace. |
# of Decoded Waveforms |
Up to 4 buses may be decoded at one time.
In addition, zooms can be displayed (with decoded information). |
Location |
Overlayed over DATA waveform, on Grid.
(Note: Use multi-grid if there is more than one decoder ON) |
Visual Aid |
Color coding for Frame, Break, Synch, ID, ID Parity, Data, CRC.
Decode information is intelligently annotated based on timebase setting. |
|
Trigger Capability |
Format |
Hexadecimal, Binary |
Trigger Setup |
Trigger on Address, Address & Data, Status, Error, Response Time, IMG (Intermessage Gap Time).
These can be further qualified by selecting the message type: BC-RT, RT-BC, RT-RT, Broadcast. |
ADDRESS (ID)
Condition Setup |
Specify one terminal address — all are 5 bits
Specify one subaddress — all are 5 bits
Specify word count/mode code — all are 5 bits
Selection for Transmit/Receive bit as 0/1/X |
DATA Condition Setup |
<=, <, =, >, >=, <>, in range, out of range, don’t care. |
DATA Setup |
Hexadecimal: # Data Bytes = 0 to 16. Data can be defined by nibble.
Triggers on that data pattern regardless of position or in user settable location.
Binary: Any combination of 0,1, or X for 1-64 bits.
Triggers on that data pattern regardless of position or in user settable location. |
Status Setup |
Status triggers indicated by a check box.
Selection of status triggers are: Message Error, Instrumentation, Service Request, Broadcast
Command Received, Busy, Subsystem Flag, Dynamic Bus Acceptance, Terminal Flag. |
Error Setup |
Selection of error triggers indicated by a check box. Seletion of errors listed below. |
Response Time
Trigger Setup |
<=, <, =, >, >=, <>, in range, out of range, don’t care. |
Intermessage Gap Setup |
<=, <, =, >, >=, <>, in range, out of range, don’t care. |
ACK Condition Setup |
Not applicable. |
Bit Rates |
Any.
Auto-detected. |
Trigger Input |
Any analog Channel or the EXT input. |
Trigger Design |
Internal to oscilloscope, settable like any other oscilloscope trigger. |
|
Search Capability |
Pattern Search |
Search by Index, Time, Message, Transfer, Type, Summary, Sync, RT Address, Transmit/Receive
Flag, Subaddress, Count, Mode Code, Parity, Response Time, RT Address Ack, Message Error,
Instrumentation, Service Request Bits, Reserved Bits, Broadcast Received, Busy, Subsystem
Flag, Dynamic Bus Control Accepted, Terminal Flag, Data, Intermessage Gap Time, Status, and
Attributes Fields. |
|
Other |
Compatible With... |
Fully compatible with WaveMaster®8 Zi, WavePro®7 Zi, WaveRunner®Xi/Xi-A, and
WaveSurfer®Xs/Xs-A. |