POWER SUPPLY |
100V / 120V / 220V / 240V. |
LINE FREQUENCY |
50Hz to 60Hz and 400Hz, automatically sensed at power-up. |
POWER CONSUMPTION |
40 VA. |
OPERATING ENVIRONMENT |
Specified for 0°C to 50°C. Specified to 80% R.H. at 35°C. RH at 35°C and at an altitude of up to 2,000 meters. |
STORAGE ENVIRONMENT |
–40°C to 70°C. |
WARRANTY |
3 years. |
SAFETY |
Conforms to European Union Directive 73/23/EEC EN61010-1. |
EMC |
Complies with European Union Directive 89/336/EEC, EN61326-1. |
VIBRATION |
MIL-PRF-28800F Class 3 Random. |
WARMUP |
1 hour to rated accuracy. |
DIMENSIONS |
Rack Mounting: 89mm high × 213mm wide × 370mm deep (3 1 / 2 in × 8 3 / 8 in × 14 9 / 16 in). |
Bench Configuration (with handle and feet) |
104mm high × 238mm wide × 370mm deep (4 1 / 8 in × 9 3 / 8 in × 14 9 / 16 in). |
NET WEIGHT |
4.2kg (8.8 lbs). |
SHIPPING WEIGHT |
5kg (11 lbs). |
VOLT HERTZ PRODUCT |
=8 × 10 7 V·Hz. |
ACCESSORIES SUPPLIED |
Model 1751 Safety Test Leads, User Manual, Service Manual. |
VOLTAGE RANGE |
100mV, 1V, 10V, 100V, 750V (user selectable). |
INPUT IMPEDANCE |
1MΩ paralleled by <100pF |
DISPLAY RANGE |
0–100% or 0–100.00dB |
RESOLUTION |
0.0001% or 0.00001dB |
FUNDAMENTAL FREQUENCY RANGE |
20Hz–20kHz |
HARMONIC FREQUENCY RANGE |
40Hz–50kHz |
FREQUENCY RESOLUTION |
0.008Hz |
FREQUENCY ACCURACY |
±0.01% of reading |
FREQUENCY TEMPERATURE COEFFICIENT |
=100ppm over operating temperature range |
MEASUREMENT MODE |
ACCURACY (1 Year, 23°C ±5°C) |
RESIDUAL DISTORTION |
THD and individual harmonic magnitudes |
±0.8dB, 20Hz to 20kHz 2 |
0.004% or –87dB
20Hz to 20kHz |
THD + n |
±1.5 dB,
100Hz to 20kHz 2 |
0.056% or –65dB
20Hz to 20kHz |
SINAD |
±1.5dB
100Hz to 20kHz |
+65dB
20Hz to 20kHz |
AC Level
V rms |
±(0.13% of reading + 0.009% of range)
20Hz to 20kHz |
External Interface |
None |
C-Message |
CCITT Weighting |
CCIR/ARM |
CCIR |
“A” Weighting |
NUMBER OF HARMONICS INCLUDED IN THD CALCULATION |
2 to 64 (user selectable) |
HI AND LO CUTOFF FILTERS (bus settable) |
20Hz–50kHz. Can be combined to form brickwall bandpass filter |
Distortion Measurement Reading Rate |
FUNDAMENTAL FREQUENCY ACQUISITION MODE |
FUNDAMENTAL FREQUENCY RANGE |
MINIMUM READINGS PER SECOND |
Single acquisition or stored value |
20 Hz to 100 Hz
100 Hz to 1 kHz
1 kHz to 20 kHz |
14
24
28 |
Automatic |
20 Hz to 30 Hz
30 Hz to 400 Hz
400 Hz to 20 kHz |
5.5
6
6.6 |
Frequency Sweep Reading Rate |
NUMBER OF FREQUENCIES |
TIME (seconds) |
5 |
0.2 |
30 |
1.1 |
100 |
3.5 |
200 |
6.9 |
Notes |
- Input signal at full scale.
- V IN =20% of range and harmonics >–65dB.
- Speeds are for default operating conditions (*RST), and display off, auto range off, binary data transfer, trig delay = 0.
- Typical times: frequencies in 400–4kHz range, binary data transfer, TRIG DELAY = 0, Display OFF, Auto Range OFF. Data returned is THD measurement plus AC voltage.
|
FREQUENCY RANGE |
10–20kHz |
FREQUENCY RESOLUTION |
0.007Hz. |
FREQUENCY ACCURACY |
±(0.015% of reading + 0.007Hz) |
FREQUENCY TEMPERATURE COEFFICIENT |
<100ppm over operating temperature range. |
SOURCE OUTPUT |
Waveform |
Sinewave |
Amplitude Range |
2V rms (50Ω and 600Ω) or 4V rms (HI Z) |
Amplitude Resolution |
0.5mV rms (50Ω and 600Ω) or 1mV rms (HI Z) |
Amplitude Accuracy |
±(0.3% of setting + 2mV) |
Amplitude Temperature Coefficient |
Typically 0.015%/°C |
Amplitude Flatness |
±0.1dB |
Output Impedance |
50O ± 1Ω or 600O ± 10Ω, user selectable |
THD |
–64dB |
Noise |
100µV rms |
DC Offset Voltage |
±1.2mV |
INV/PULSE OUTPUT (SINEWAVE MODE) |
Frequency |
Same as source output |
Amplitude Range |
2V rms (50Ω and 600Ω) or 4V rms (HI Z) |
Amplitude Resolution |
0.5mV (50Ω and 600Ω) or 1mV rms (HI Z) |
Amplitude Accuracy |
±(2.0% of setting + 2mV) |
Amplitude Flatness |
±0.1dB |
Output Impedance |
Same as Source Output setting |
THD |
–64dB |
Noise |
100µV rms |
DC Offset Voltage |
±1.1mV typ., ±13mV max |
INV/PULSE OUTPUT (PULSE MODE) |
Frequency |
Same as source output |
Duty Cycle |
45% ±3% |
Output Impedance |
Same output impedance as the source output |
Amplitude |
0.0V ±0.07V to 4.9V ±0.12V pulse open circuit
0.0V ±0.05V to 3.3V ±0.08V pulse 100Ω load |
Overshoot |
1.0V maximum pulse open circuit
0.2V maximum with 100Ω load pulse open circuit |
Undershoot |
1.1V maximum pulse open circuit
0.45V maximum with 100Ω load pulse open circuit |
Notes |
- 1 year, 23°C ±5°C
- Measured at V OUT = 0V with gain 100 amplifier and 2-pole 50kHz low pass filter, Inv/Pulse in sinewave mode, HI Z output impedance, and no load
- With HI Z output impedance and 1m 50Ω coaxial cable
- HI Z output impedance, no load
- 4V output
- THD measurement includes harmonics 2 through 5, 1V rms output, HI Z, no load.
|
CONDITIONS: MED (1 PLC)1 OR SLOW (10 PLC)ACCURACY: ±(PPM OF READING + PPM OF RANGE) OR MED (1 PLC) WITH FILTER OF 10(PPM = PARTS PER MILLION) (E.G., 10PPM = 0.001%) |
FUNCTION |
RANGE |
RESOLUTION |
TEST CURRENT
OR BURDEN
VOLTAGE (±5%) |
INPUT
RESISTANCE |
24 HOUR
23°C ± 1° |
90 DAY
23°C ± 5° |
1 YEAR
23°C ± 5° |
TEMPERATURE
COEFFICIENT
0°–18°C & 28°–50°C |
VOLTAGE |
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V |
0.1 µV
1.0 µV
10 µV
100 µV
1 mV |
|
> 10 GΩ
> 10 GΩ
> 10 GΩ
10 MΩ ±1%
10 MΩ ±1% |
30 + 30
15 + 6
15 + 4
15 + 6
20 + 6 |
40+35
25+7
20+5
30+6
35+6 |
50+35
30+7
30+5
45+6
45+6 |
52 + 6
2 + 1
2 + 1
5 + 1
5 + 1 |
RESISTANCE |
1100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ |
100 µΩ
1 mΩ
10 mΩ
100 mΩ
1 Ω
10 Ω
100 Ω |
1 mA
1 mA
100 µA
10 µA
10 µA
700 nA // 10MΩ
700 nA // 10MΩ |
|
30 + 30
20 + 6
20 + 6
20 + 6
20 + 6
300 + 6
1600 + 30 |
80 + 40
80 + 10
80 + 10
80 + 10
80 + 10
450 + 10
2000 + 30 |
100 + 40
100 + 10
100 + 10
100 + 10
100 + 10
600 + 10
2200 + 30 |
8 + 6
8 + 1
8 + 1
8 + 1
8 + 1
70 + 1
385 + 1 |
CURRENT |
10.00000 mA
100.0000 mA
1.000000 A
3.00000 A |
10 nA
100 nA
1 µA
10 µA |
< 0.15 V
< 0.03 V
< 0.3 V
< 1 V |
|
60 + 30
100+300
200+30
1000+15 |
300+80
300+800
500 + 80
1200 + 40 |
500 + 80
500 + 800
800 + 80
1200 + 40 |
50 + 5
50 + 50
50 + 5
50 + 5 |
CONTINUITY 2W |
1 kΩ |
100 mΩ |
1 mA |
|
40 + 100 |
100 + 100 |
120 + 100 |
8 + 1 |
DIODE TEST |
3.00000 V
10.00000 V
10.00000 V |
10 µV
10 µV
10 µV |
1 mA
100 µA
10 µA |
|
20 + 6
20 + 6
20 + 6 |
30 + 7
30 + 7
30 + 7 |
40 + 7
40 + 7
40 + 7 |
8 + 1
8 + 1
8 + 1 |
FUNCTION |
DIGITS |
READINGS/s |
PLCs |
DCV (all ranges),
DCI (all ranges), and
2W Ohms (<10M range) |
6 1 / 2
6 1 / 2
6 1 / 2
5 1 / 2
5 1 / 2
5 1 / 2
4 1 / 2 |
5
30
50
270
500
1000
2000 |
10
1
1
0.1
0.1
0.04
0.01 |
RANGE CHANGE |
50 / s |
FUNCTION CHANGE |
45 / s |
AUTORANGE TIME |
<30 ms |
ASCII READINGS TO RS-232 (19.2K BAUD) |
55 / s. |
MAX. INTERNAL TRIGGER RATE |
2000 / s. |
MAX. EXTERNAL TRIGGER RATE |
400 / s. |
LINEARITY OF 10VDC RANGE |
±(2ppm of reading + 1ppm of range) |
DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION |
1000V, all ranges |
MAXIMUM 4WΩ LEAD RESISTANCE |
10% of range per lead for 100Ω and 1kΩ ranges; 1kΩ per lead for all other ranges |
DC CURRENT INPUT PROTECTION |
3A, 250V fuse |
SHUNT RESISTOR |
0.1Ω for 3A, 1A and 100mA ranges. 10Ω for 10mA range |
CONTINUITY THRESHOLD |
Adjustable 1O to 1000Ω. |
AUTOZERO OFF ERROR |
Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change |
OVERRANGE |
120% of range except on 1000V, 3A and Diode. |
- Add the following to ppm of range accuracy specification based on range:1V and 100V, 2ppm; 100mV, 15ppm; 100Ω, 15ppm; <1MΩ, 2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm
- Speeds are for 60 Hz operation using factory default operating conditions (*RST). Autorange off, Display off, Trigger delay = 0
- Speeds include measurement and binary data transfer out the GPIB
- Auto zero off
- Sample count = 1024, auto zero off
- Auto zero off, NPLC = 0.01
- Ohms = 24 readings/second
- 1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up
- For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
- Add 120ms for ohms
- Must have 10% matching of lead resistance in Input HI and LO
- For line frequency ±0.1%
- For 1kΩ unbalance in LO lead
- Relative to calibration accuracy
- Specifications are for 4-wire ohms. For 2-wire ohms, add 1Ω additional uncertainty.
- For rear inputs. Add the following to temperature Coefficient ‘ppm of reading’ uncertainty; 10MΩ 70ppm, 100MΩ 385ppm. Operating environment specified for 0°C to 50°C, 50% RH at 35°C
|
ACCURACY 1 : ±(% of reading + % of range), 23°C ±5 °C |
VOLTAGE |
RANGE |
CALIBRATION
CYCLE |
3 Hz–
10 Hz |
10 Hz–
20 kHz |
20 kHz–
50 kHz |
50 kHz–
100 kHz |
100 kHz–
300 kHz |
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
750.000 V |
0.1 µV
1.0 µV
10 µV
100 µV
1 mV |
90 Days
1 Year |
0.35+0.03
0.35+0.03 |
0.05+0.03
0.06+0.03 |
0.11+0.05
0.12+0.05 |
0.60+0.08
0.60+0.08 |
4 + 0.5
4 + 0.5 |
|
TEMPERATURE
COEFFICIENT/°C |
0.035+0.003 |
0.005+0.003 |
0.006+0.005 |
0.01+0.006 |
0.03+0.01 |
CURRENT
RANGE |
RESOLUTION |
CALIBRATION
CYCLE |
3 Hz–
10 Hz |
10 Hz–
3 kHz |
3 kHz-
5 kHz |
1.000000 A |
1 µA |
90 Day/1 Year |
0.30+0.04 |
0.10+0.04 |
0.14+0.04 |
3.00000 A |
10 µA |
90 Day/1 Year |
0.35+0.06 |
0.15+0.06 |
0.18+0.06 |
|
TEMPERATURE
COEFFICIENT/°C |
0.035+0.006 |
0.015+0.006 |
0.015+0.006 |
CREST FACTOR |
1–2 |
2–3 |
3–4 |
4–5 |
ADDITIONAL ERROR |
0.05 |
0.15 |
0.30 |
0.40 |
FUNCTION |
DIGITS |
READINGS/s |
RATE |
BANDWIDTH |
ACV (all ranges) |
6 1 / 2 |
2s/reading |
SLOW |
3 Hz–300 kHz |
ACI (all ranges) |
6 1 / 2 |
1.4 |
MED |
30 Hz–300 kHz |
|
6 1 / 2 |
4.8 |
MED |
30 Hz–300 kHz |
|
6 1 / 2 |
2.2 |
FAST |
300 Hz–300 kHz |
|
6 1 / 2 |
35 |
FAST |
300 Hz–300 kHz |
FUNCTION |
SLOW |
MED |
FAST |
20Hz – 30Hz |
0 |
0.3 |
- |
30Hz – 50Hz |
0 |
0 |
- |
50Hz – 100Hz |
0 |
0 |
1.0 |
100Hz – 200Hz |
0 |
0 |
0.18 |
200Hz – 300Hz |
0 |
0 |
0.10 |
> 300Hz |
0 |
0 |
0 |
FUNCTION/RANGE CHANGE |
4 / s |
AUTORANGE TIME |
<3 s |
ASCII READINGS TO RS-232 (19.2K BAUD) |
50 / s |
MAX. INTERNAL TRIGGER RATE |
300 / s |
MAX. EXTERNAL TRIGGER RATE |
260 / s |
INPUT IMPEDANCE |
1MΩ ±2% paralleled by <100pF |
ACV INPUT PROTECTION |
1000Vp |
MAXIMUM DCV |
400V on any ACV range |
ACI INPUT PROTECTION |
3A, 250V fuse |
BURDEN VOLTAGE |
1A Range: <0.3V rms. 3A Range: <1V rms |
SHUNT RESISTOR |
0.1Ω on all ACI ranges |
AC CMRR |
>70dB with 1kΩ in LΩ lead |
MAXIMUM CREST FACTOR |
5 at full scale |
VOLT HERTZ PRODUCT |
=8 × 10 7 V·Hz |
OVERRANGE |
120% of range except on 750V and 3A ranges |
- Specifications are for SLOW rate and sinewave inputs >5% of range
- Speeds are for 60 Hz operation using factory default operating conditions (*RST). Auto zero off, Auto range off, Display off, includes measurement and binary data transfer out the GPIB
- 0.01% of step settling error. Trigger delay = 400ms
- Trigger delay = 0
- DETector:BANDwidth 300, NPLC = 0.01
- Maximum useful limit with trigger delay = 175ms
- Applies to non-sine waves >5Hz and <500Hz. (Guaranteed by design for Crest Factors >4.3)
- Applies to 0°–18°C and 28°–50°C
- For signal levels >2.2A, add additional 0.4% to “of reading” uncertainty
- Typical uncertainties. Typical represents two sigma or 95% of manufactured units measure < 0.35% of reading and three sigma or 99.7% < 1.06% of reading.
|
READING HOLD SENSITIVITY |
0.01%, 0.1%, 1%, or 10% of reading |
TRIGGER DELAY |
0 to 99 hrs (1ms step size) |
EXTERNAL TRIGGER LATENCY |
200µs + <300µs jitter with autozero off, trigger delay = 0 |
MEMORY |
1024 readings |
Rel, Min/Max/Average/StdDev (of stored reading), dB, dBm, Limit Test, %, and mX+b with user defined units displayed. |
dBm REFERENCE RESISTANCES |
1 to 9999O in 1Ω increments |
SCPI (Standard Commands for Programmable Instruments) |
GPIB (IEEE-488.1, IEEE-488.2) and RS-232C |
ACV
RANGE |
FREQUENCY
RANGE |
PERIOD
RANGE |
GATE
TIME |
RESOLUTION
±(ppm of
reading) |
ACCURACY
90 DAY/1 YEAR
±(% of reading) |
100 mV
to
750 V |
3 Hz
to
500 kHz |
333 ms
to
2 µs |
1 s (SLOW)
0.1 s (MED)
10 ms (FAST) |
0.333
3.33
33.3 |
0.01
0.01
0.01 |
- Specifications are for square wave inputs only. Input signal must be >10% of ACV range. If input is <20mV on the 100mV range then the frequency must be >10Hz
- 20% overrange on all ranges except 750V range
|
TYPE |
RANGE |
RESOLUTION |
90 DAY/1 YEAR (23°C ± 5°C)
ACCURACY
Relative to
Reference Junction |
J |
–200 to+760°C |
0.001°C |
±0.5°C |
K |
–200 to + 1372°C |
0.001°C |
±0.5°C |
T |
–200 to + 400°C |
0.001°C |
±0.5°C |