Bandwidth |
DC to 100 MHz
(-3dB) |
DC to 50 MHz
(-3dB) |
Maximum input current |
Maximum input range |
30 A rms |
15 A rms |
Maximum peak current |
50 A peak, non-continuous |
30 A peak, non-continuous
50 A peak at pulse width <= 10 µs |
Output sensitivity
Sensitivity accuracy
(For probes only)
*reading: output voltage
|
0.1 V/A
(DC, 45 to 66 Hz, for maximum continuous input current range)
•0 to 30 A rms
±1.0% (of reading) ±1 mV, at 23 ±3°C
•30 A rms to 50 A peak
±2.0% (of reading), at 23 ±3°C |
0.1 V/A
(DC, 45 to 66 Hz, for maximum continuous input current range)
•At 23 ±3°C
±0.5% (of reading) ±1 mV |
Temperature characteristics of sensitivity |
Within ±2% (in the range from 0 to 40°C) |
Noise |
2.5 mA rms or less (observation using an oscilloscope with a 20 MHz band) |
Power consumption |
5.3 VA max. |
3 VA max. |
Power supply voltage |
±12 V ±0.5 V |
±12 V ±1 V |
Operating temperature and humidity |
0 to +40°C, 80% RH or less
(no condensation) |
Storage temperature and humidity |
-10 to +50°C, 80% RH or less
(no condensation) |
Operating elevation |
2000 m or less (air pressure approx. 79 kPa) |
Maximum usable circuit voltage |
300 V, CAT I (insulating conductor) |
Conductor diameter that can be measured |
ø 5 mm |
Cable length |
Sensor cable |
Approx. 1.5 m |
Power cable |
Approx. 1 m |
Size |
Sensor part |
Approx. 175 (W) x 18 (H) x 40 (D) mm |
Termination part |
Approx. 27 (W) x 55 (H) x 18 (D) mm |
Weight |
Approx. 240 g |
Approx. 230 g |
Accessories |
Instruction manual,
hard case |