Voltage Range: 0.100kV ~ 5.000kV Voltage Resolution Output: 5V/step Voltage Regulation (Line and Load): 1% + 5 V Voltage Accuracy: ±1% of reading ±10V Current Sourcing: 30 ~ 40mA(above 500V, test time max:180sec) 0.10 ~ 29.99mA(above 500V,continuous test) 0.10 ~ 10mA(below 500V, continuous test) Current Limit: 0.10 ~ 40mA, 0.02mA/step Current Accuracy: ±1% of reading ±80µA Voltage Range: 0.100kV ~ 6.000kV Voltage Resolution Output: 5V/step Voltage Regulation (Line and Load): 1% + 5 V Voltage Accuracy: ±1% of reading ±10V Current Sourcing: 0.10 ~ 10.00mA(above 500V,continuous test) 0.10 ~ 2mA( below 500V,continuous test) 0.10 ~ 10mA, 0.01mA/step Current Accuracy: ±1% of reading ±50µA DC Voltage: 50V/100V/500V/1000V Resistance Accuracy (50V/100V): 1~50MO: ±5% of reading ±1count 51~200MO: ±10% of reading ±1count 201~1990MO: ±20% of reading ±1count Resistance Accuracy (500V/1000V): 1~500MO: ±5% of reading 1±count 501~2000MO: ±10% of reading ±1count 2001~9990MO: ±20% of reading ±1count Output Current Range: 0.1A Compliance Voltage Range: 10% Current Resolution: 0.1O @ 1O Detect Current: 40 level (1 ~ 40mA) Groups: 10 Steps Per Group: 16 Remote I/O: Standard RS-232C: Standard Scan: Standard GPIB: Option Real Panel Output: Option Scanner Box: Option(HSB-001-1: 8ch HV + 8ch GB HSB-001-2: 16 ch HV) |