Frequency Range |
150kHz ~ 1000MHz |
Frequency Resolution |
1kHz C.F. entry, 40Hz Sweep resolution at 2kHz/DIV |
Frequency Display |
6 1/2 digit setting |
Frequency Control |
Digital phase locked |
Frequency Stability |
±2ppm/year aging,±10ppm, 0 to 50°C |
Frequency Spans |
Zero, 2kHz to 100MHz/DIV, in a 1-2-5 sequence |
BANDWIDTH |
Resolution Bandwidths |
3kHz,30kHz, 220kHz, 4MHz |
Resolution BW Accuracy |
15% |
Video Bandwidth |
1.6kHz/90kHz coupled with RBW |
AMPLITUDE |
Reference Level Range |
-30dBm to +20dBm |
Reference Level Accuracy |
±1dB at 80MHz |
Input Level Range |
-100dBm to +20dBm |
Noise Floor |
-95dBm@30kHz RBW, -100dBm typical/-75dBm:150kHz ~ 10MHz |
Amplitude Display Range |
75dB |
Amplitude Accuracy |
±1.5dB typical@0dBm, 80MHz |
Amplitude Level Linearity |
±1.5dB over 70dB |
Ref. Level Freq. Flatness |
±1.5dB over 100MHz, ±2.5dB typical over entired band/±3dB:150kHz ~10MHz |
Harmonic Spur Response |
<-40dBc, RF input < selected reference |
Non-Harmonic Spur Response |
<-60dBc typical down from reference level, average, 5MHz/div |
Intermodulation(3rd) |
<-70dBc, @-40dBm input, 2 tones, 2MHz apart/<-45dBc:150kHz ~10MHz |
Phase Noise |
-77dBc/Hz at 1GHz, 30kHz offset |
INPUT |
Input Overload Protection |
+30dBm continuous, ±25VDC |
Impedance |
50 ohm nominal |
Return Loss |
<16 dBRL (VSWR < 1.35) |
Input Attenuation |
50dB to 0dB in 10 dB steps coupled to reference level |
Connector |
Type N female |
MARKER |
Number of Markers |
2 |
Marker Resolution |
0.1dB, 1kHz |
Marker Mode |
Absolute, Relative, PK -> Marker, Marker -> Center |
Marker Accuracy |
0.1dB±amplitude accuracy |
FUNCTIONS |
Memory |
9 memories of save/recall |
Trace |
Max. Hold, Average(2 ~ 32 traces), Freeze(Hold) |
Setup |
Access parameters |
FM/AM DEMODULATOR |
WB FM |
120kHz deviation |
MB FM |
75kHz deviation |
NB FM |
30kHz deviation |
AM Outputs |
Internal speaker, 3.5mm stereo jack, wired for mono operation |
REMOTE DISPLAY SOFTWARE & RS-232 |
Connecting to PC and getting trace from GSP-810, software will be downloaded from here |
GENERAL |
Power Source |
100/120/220/230V AC±10%, 50/60Hz |
Accessory |
Operation manual x 1 |
Dimensions and Weight |
310(W) x 150(H) x 455(D)mm, approx. 8.5kg |
OPTIONS |
TRACKING GENERATOR (OPTION 01) |
Frequency Range |
10MHz to 1000MHz |
Amplitude Range |
0 to -50dBm |
Amplitude Resolution |
1dB |
Amplitude Accuracy |
±1dB@0dBm, 80MHz |
Attenuation Accuracy |
±1dB@, 50MHz |
Amplitude Flatness |
±1dB@, 10MHz/DIV, ±1.5dB@0dB,entire band |
Harmonics |
<-30dBc |
Reverse Power |
<+30dBm |
Impedance |
50 ohm nominal |
Return Loss |
<10 dBRL (VSWR<2) |
Connector |
Type N female |
POWER METER (OPTION 02) |
Frequency Range |
10MHz to 2GHz, usable to 2.7GHz |
Power Level Range |
-20dBm to +23dBm, usable to +30dBm |
Power Level Overload |
+40dBm<10% duty cycle, <10mS duration |
Return Loss |
<1:1.35 VSWR into 50 ohms, <1:1.25 typical |
Readout Resolution |
0.2mW, 100mW scale, 2µW, 1mW scale; 0.1dB, Log scale |
Accuracy |
±(10% rdg ±1digit) |
Readout |
mW or dBm |
REMOTE CONTROL SOFTWARE (OPTION 03) |
Connecting PC to get trace and provide the control for setting |