Input Specifications |
Measurement line type |
Single-phase 2-wire(1P2W) |
Input methods |
Voltage |
Isolated input, resistive voltage divider method |
Current |
Isolated input, shunt input method |
|
Voltage measurement ranges |
AUTO/ |
6 .0000 V |
15.000 V |
30.000 V |
60.000 V/
1.0000 kV |
150.00 V/ |
300.00 V/ |
600.00 V |
|
Current measurement ranges |
AUTO/ |
1.0000 mA |
2.0000 mA/ |
5.0000 mA |
10.000 mA |
20.000 mA/ |
50.000 mA |
100.00 mA/ |
200.00 mA/ |
500.00 mA/ |
1.0000 A/ |
2.0000 A/ |
5.0000 A/ |
10.000 A/ |
20.000 A |
|
|
Power ranges |
Depends on the combination of voltage and current ranges;
From 6.0000 mW to 20.000 kW (also applies to VA, var) |
Input resistance |
Voltage input terminal:
Approx.2 MΩ
Current input terminal:
1 mA to 100 mA range 520 mΩ or less
200 mA to 20 A range 15 mΩ or less |
Basic Measurement Specifications |
Measurement method |
Simultaneous voltage and current digital sampling, zero-cross simultaneous calculation |
Sampling frequency |
Approx. 700 kHz |
A/D converter resolution |
16-bit |
Frequency bandwidth |
DC, 0.1 Hz to 100 kHz
(Values within 0.1Hz = f < 10 Hz are for reference only) |
Synchronization source |
U, I, DC (fixed to 200 ms) |
Zero-cross Filter |
100 Hz: 0.1 Hz to 100 Hz 500 Hz: 0.1 Hz to 500 Hz
5 kHz: 0.1 Hz to 5 kHz 100 kHz: 0.1 Hz to 100 kHz |
Effect of power factor |
±0.1%f.s. or less (45 to 66 Hz, at power factor = 0)
Internal circuitry voltage/current phase difference: ±0.0573° |
Effect of common mode voltage |
±0.01%f.s. or less (600 V, 50 Hz/60 Hz, applied between input terminals and enclosure)° |
Voltage/ Current/ Active Power Measurement Specifications |
Measurement types |
Rectifiers: AC+DC, DC, AC, FND, AC+DC Um |
Effective measuring range |
Voltage
±1% to ±150% of the range
However, up to ±1500 V peak value and 1000 V RMS value
Current
±1% to ±150% of the range
Active Power
±0% to ±225% of the range.
However, valid when the voltage and current fall within the effective measurement range. |
Display range |
Voltage
Up to ±152% of the range. However, zero-suppression when less than ±0.5%
Current
Up to ±152% of the range.
However, zero-suppression when less than ±0.5% or less than ±9 µA.Active Power
±0% to ±231.04% of the range (no zero-suppression) |
Polarity |
Voltage/ Current
Displayed when using DC rectifier
Active Power
Positive : Power consumption (no polarity display)
Negative : generation or regenerated power |
Frequency Measurement Specification |
Number of measurement channels |
2 (Voltage, current) |
Measurement method |
Calculated from input waveform period (reciprocal method) |
Measurement ranges |
100 Hz/ 500 Hz/ 5 kHz/ 100 kHz (linked to zero-cross filter) |
Measurement accuracy |
±0.1% rdg. ±1 dgt. However, for 1 mA range, ±0.2% rdg. ±1 dgt. |
Effective measuring range |
0.1 Hz to 100 kHzFor sine wave input that is at least 20% of the measurement source’s measurement rangeMeasurement lower limit frequency setting: 0.1 sec. / 1 sec. / 10 sec. (linked to synchronization timeout setting) |
Display format |
0.1000 Hz to 9.9999 Hz, |
9.900 Hz to 99.999 Hz, |
99.00 Hz to 999.99 Hz, |
0.9900 kHz to 9.9999 kHz, |
9.900 kHz to 99.999 kHz, |
99.00 kHz to 100.00 kHz |
|
Apparent Power/ Reactive Power/ Power Factor/ Phase Angle Measurement Specifications |
Rectifier |
2 (Voltage, current)
Apparent Power/ Reactive Power/ Power Factor
AC+DC, AC, FND, AC+DC Umn
Phase Angle
AC, FND |
Effective measuring range |
As per voltage, current, and active power effective measurement ranges |
Display range |
Apparent Power/ Reactive Power
0% to 231.04% of the range (no zero-suppression)
Power Factor
±0.0000 to ±1.0000
Phase Angle
+180.00 to -180.00 |
Polarity |
Reactive Power/ Power Factor/ Phase Angle
Polarity is assigned according to the lead/lag relationship of the voltage waveform rising edge and the current waveform rising edge.
+: When current lags voltage (no polarity display)
-: When current leads voltage |
Voltage Waveform Peak Value/ Current Waveform Peak Value Measurement Specification |
Measurement method |
Measures the voltage waveform’s peak value (for both positive and negative polarity) based on sampled instantaneous voltage values. |
Range configuration |
Voltage
Voltage range |
Voltage peak range |
6.0000 V |
36.000 V |
15.000 V |
90.000 V |
30.000 V |
180.00 V |
60.000 V |
360.00 V |
150.00 V |
900.00 V |
300.00 V |
1.8000 kV |
600.00 V |
3.6000 kV |
1.0000 kV |
6.0000 kV |
Current
Current range |
Current peak range |
1.0000 mA |
6.0000 mA |
2.0000 mA |
12.000 mA |
5.0000 mA |
30.000 mA |
10.000 mA |
60.000 mA |
20.000 mA |
120.00 mA |
50.000 mA |
300.00 mA |
100.00 mA |
600.00 mA |
200.00 mA |
1.2000 A |
500.00 mA |
3.0000 A |
1.0000 A |
6.0000 A |
2.0000 A |
12.000 A |
5.0000 A |
30.000 A |
10.000 A |
60.000 A |
20.000 A |
120.00 A |
|
Measurement accuracy |
±2.0%f.s. at DC and when 10 Hz = f = 1 kHz (f.s.: current peak range).
Provided as reference value when 0.1 Hz = f < 10 Hz and when 1 kHz < f.
The above measurement accuracy is multiplied by 2 for the 1 mA range. |
Effective measuring range |
±5% to ±100% of current peak range, however, up to ±60 A |
Display range |
Up to ±102% of current peak range, however, the value 0 will be displayed if the current RMS value triggers the instrument's zero suppression function. |
Voltage Crest Factor/Current Crest Factor Measurement Specifications |
Measurement method |
Calculates the ratio of the voltage waveform peak value to the voltage RMS value. |
Effective measuring range |
As per voltage and voltage waveform peak value, or current and current waveform peak value effective measurement ranges |
Display range |
1.0000 to 612.00 (no polarity) |
Voltage Ripple Rate/ Current Ripple Rate Measurement Specifications |
Measurement method |
Calculates the AC component (peak to peak [peak width]) as a proportion of the voltage or current DC component. |
Effective measuring range |
As per voltage and voltage waveform peak value, or current and current waveform peak value effective measurement ranges. |
Display range |
0.00 to 500.00 (No polarity) |
Maximum Current Ratio Measurement Specifications (MCR) |
Measurement method |
Calculates the ratio of the current crest factor to the power factor. (MCR) = (Current Crest Factor) / (Power Factor) |
Effective measuring range |
As per power factor (voltage, current, active power) and current crest factor (current, current waveform peak value) effective measurement ranges. |
Display range |
1.0000 to 6.1200 M (no polarity) |
Synchronized control |
Functions |
The timing of calculations; display updates; data updates; integration start, stop, and reset events; display hold operation; key lock operation; and zero-adjustment operation for the slave PW3335 series is synchronized with the master PW3335 series. Synchronization with the PW3336 series and PW3337 series is also supported. |
Terminal |
BNC terminal × 1 (non-isolated) |
Terminal name |
External synchronization terminal (EXT.SYNC) |
I/O settings |
Synchronized control function off (signals input to the external synchronization terminal (EXT.SYNC) are ignored) |
I/O settings |
Off
Synchronized control function off (signals input to the external synchronization terminal (EXT.SYNC) are ignored)
In
The external synchronization terminal (EXT.SYNC) is set to input, and a dedicated synchronization signal can be input (slave).
Out
The external synchronization terminal (EXT.SYNC) is set to output, and a dedicated synchronization signal can be output (master). |
Number of units for which synchronized control can be performed |
Up to 7 slaves per master
(total of 8 units including the PW3336/PW3337 series) |
Display Specifications |
Display |
7-segment LED |
Number of display parameters |
4 (display area a, b, c, and d) |
Display resolution |
Other than integrated values: 99999 count (5 digits) Integrated values: 999999 count (6 digits) |
Display update rate |
200 ms ±50 ms (approx. 5 updates per sec.) to 20 s (varies with number of averaging iterations setting) |
GP-IB interface (PW3335-01 and PW3335-04) |
Method |
Compliant with IEEE488.1 1987, in reference to IEEE488.2 1987
Interface functions
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0 |
Address |
00 to 30 |
RS-232C interface (PW3335, PW3335-02, PW3335-03, and PW3335-04) |
Connector |
D-sub 9-pin connector × 1 |
Communication method |
Full duplex, Start-stop synchronization
Stop bits: 1 (fixed)
Data length: 8 (fixed)
Parity: None |
Communication speed |
9600 bps/ 38400 bps |
LAN interface |
Connector |
RJ-45 connector × 1 |
Electrical specifications |
Compliant with IEEE802.3 |
Transmission method |
10Base-T/ 100Base-TX (automatic detection) |
Protocol |
TCP/ IP |
Functions |
HTTP server (remote operation, firmware updates)
Dedicated ports (command control, data transfer)
Remote control by controller |
General Specifications |
Product warranty period |
1 year |
Operating environment |
Indoors, altitude up to 2000 m (6562 ft.), pollution degree 2 |
Operating temperature and humidity |
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation) |
Storage temperature and humidity |
-10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation) |
Dielectric strength |
4290 V rms AC (current sensitivity: 1 mA)
Between the voltage input terminals and a connection consisting of chassis, interfaces, and output terminals
Between the current input terminals and a connection consisting of chassis, interfaces, and output terminals
Between the voltage input terminals and current input terminals |
Maximum rated voltage to earth |
Voltage input terminal, Current input terminal
Measurement category III 600 V (anticipated transient overvoltage: 6000 V)
Measurement category II 1000 V (anticipated transient overvoltage: 6000 V) |
Maximum input voltage |
Between the voltage input terminals U and ±
1000 V, ±1500 V peak |
Maximum input voltage |
Between the voltage input terminals U and ±
1000 V, ±1500 V peak |
Maximum input current |
Between the current input terminals I and ±
200 mA to 20 A range 30 A, ±100 A peak
1 mA to 100 mA range 20 A, ±30 A peak |
Applicable Standards |
Safety |
EN61010 |
EMC |
EN61326 Class A
EN61000-3-2
EN61000-3-3 |
|
Rated supply voltage |
100 V AC to 240 V AC
50 Hz/60 Hz |
Maximum rated power |
30 VA or less |
Dimensions |
Approx. 210W × 100H × 245D mm (8.27”W × 3.94”H × 9.65”D)
(excluding protrusions) |
Mass |
Approx. 3 kg (105.8 oz.) |
Accessories |
Instruction manual ×1
Power cord ×1
Voltage and current input terminal safety cover ×2 |