Supported instrument |
[Used as standalone product (Data collected manually) ]
Windows PC or Windows tablet (CD-R with software included)
Android smartphone or Android tablet terminal
(Software can be downloaded free of charge from Google Play.)
*Communication range varies with the performance of the computer or tablet (up to a line-of-sight distance of roughly 30 m)
[Used as logging module (Real-time measurement) ]
Device can be used as an LR8410 logging module to record and display data in real time and to control up to 7 units, Communication distance: 30 m |
Number of channels |
2 ch (isolated; select voltage of thermocouple for each channel), Input terminals: M3 screw type terminal block |
Measurement items |
Voltage/ Thermocouple (K, T) |
Maximum input voltage |
DC ±50 V, Max. inter-channel voltage DC 70 V |
Measurement range |
[Voltage] ±50 mV to ±50 V , Max. resolution 0.01 mV
[Thermocouple] -200 ˚C to 999.9 ˚C, Thermocouples (K, T), Max. resolution 0.1 ˚C |
Measurement accuracy |
[Voltage] ±0.05 mV (50 mV range)
[Thermocouple] ±0.8 ˚C (Thermocouple K -100 ˚C to 999.9 ˚C)
*Reference junction compensation: Switchable between internal and external
*Reference junction compensation accuracy: ±0.5 ˚C
(When using internal compensation, add to thermocouple measurement accuracy.)
*Temperature characteristics: Add (measurement accuracy × 0.1) / ˚C to measurement accuracy. |
Display items |
Measurement value, date, time, number of recorded data, maximum value, minimum value, and average value |
Functions |
Alarm, Scaling, Recording operation hold function, Erroneous operation prevention, Comment recording function, Power saving function, Authentication function |
Recording |
[Capacity] 500,000 data items for each channel [Mode] Instantaneous value [Interval] 0.1 to 30 sec, 1 to 60 min, 16 selections |
Power source |
AC ADAPTER Z2003 (AC100 V to 240 V, 50 Hz/60 Hz), AA alkaline batteries (LR6) ×2, External power DC5 V to 13.5 V (can also be supplied from USB bus power, with a conversion cable) |
Continuous operating time ([Capacity] 500,000 data items for each channel) (23˚C) |
2.5 months (Recording interval of 1 min, Bluetooth® OFF)
7 days (Recording interval of 1 sec, Bluetooth® ON)
2 days (Recording interval of 0.1 sec, during real-time measurement with the LR8410) |
Dimensions, Weight |
85W×75H×38D mm (3.35W×2.95H×1.50D in), 126 g (Not including the battery) |
Accessories |
CD-R ×1 (Instruction Manual, Logger Utility, Wireless Logger Collector) , Measurement Guide ×1, Caution for Using Radio Waves × 1, AA alkaline batteries (LR6) ×2 |