Accuracy |
±2°C (±3.6°F) or ±2% of reading, for ambient temperature 15°C–35°C (59°F–95°F) and object temperature above 0°C (32°F) |
Alarm Functions |
On any selected measurement function, digital in, and internal camera temperature |
Alarm Output |
• Digital out
• E-mail (SMTP) (push)
• EtherNet/IP (pull)
• File transfer (FTP) (push)
• Modbus TCP slave (pull)
• MQTT (push)
• Query over RESTful API (pull)
• Store image or video |
Base Mounting |
4× M4 on 4 sides |
Encoding |
H.264, MPEG4, or MJPEG |
Housing material |
Aluminium |
IR Resolution |
320 × 240 |
Object Temperature Range |
-20°C to 120°C (-4°F to 248°F), 0°C to 650°C (32°F to 1202°F), 300°C to 2000°C (572°F to 3632°F) |
Size (L x W x H) |
123 × 77 × 77 mm (4.84 × 3.03 × 3.03 in) |
Thermal Sensitivity/NETD |
Lens dependent |
Web interface |
Yes |
Weight |
0.82 kg (1.8 lb), camera only |
Contrast Enhancement |
FSX® / Histogram equalization (IR only) |
Multi-Streaming |
Yes |
FLIR Screen-EST Mode |
No; only offered when purchased with the Advanced configuration. |
Compliance & Certifications |
ONVIF Conformance |
No |
Connections & Communications |
Digital I/O connector type |
M12 12-pin A-coded, Male (shared with external power) |
Digital I/O isolation voltage |
500 VRMS |
Digital input purpose |
NUC, NUC disable, alarm |
Digital Inputs |
2x opto-isolated, Vin(low)= 0–1.5 V, Vin(high)= 3–25 V |
Digital Output Purpose |
• As a function of alarm, output to external device
• Fault (NC) |
Digital Outputs |
3x opto-isolated, 0–48 V DC, max. 350 mA; solid-state opto relay; 1x dedicated as Fault output (NC) |
Ethernet |
For control, result, image, and power |
Ethernet Communication |
TCP/IP socket-based FLIR proprietary |
Ethernet Connector Type |
M12 8-pin X-coded, female; RP-SMA, female |
Ethernet Power |
Power over Ethernet, PoE IEEE 802.3af class 3 |
Ethernet Protocols |
• EtherNet/IP
• IEEE 1588
• Modbus TCP Slave
• MQTT
• SNMP
• TCP, UDP, SNTP, RTSP, RTP, HTTP, HTTPS, ICMP, IGMP, sftp (server), FTP (client), SMTP, DHCP, MDNS (Bonjour), uPnP |
Ethernet Standard |
IEEE 802.3 |
Ethernet Type |
1000 Mbps |
RS-232 |
NA |
WiFi |
Optional feature purchased separately. RP-SMA, female connector |
Ethernet Interface |
Wired, Wi-Fi (optional) |
Multicast |
Yes |
Environmental |
Atmospheric transmission correction |
Based on inputs of distance, atmospheric temperature, and relative humidity |
EMC |
• ETSI EN 301 489-1 (radio)
• ETSI EN 301 489-17 (radio)
• EN 61000-4-8 (magnetic field)
• FCC 47 CFR Part 15 Class B (emission US)
• ISO 13766-1 (EMC - Earth-moving and building construction machinery)
• EN ISO 14982 (EMC - Agricultural and forestry machinery) |
Emissivity Correction |
Variable from 0.01 to 1.0 |
Encapsulation |
IEC 60529, IP 54, IP66 with accessory |
Humidity (Operating and Storage) |
IEC 60068-2-30/24 hours, 95% relative humidity, 25°C–40°C (77°F–104°F)/2 cycles |
Operating Temperature Range |
• -20°C to 50°C (–4°F to 122°F):
• -20°C to 40°C (-4°F to 104°F) (in free air)
• 40°C to 50°C (104°F to 122°F) (mounted on cooling plate accessory) Maximum camera case temperature: 65°C (149°F) |
Radio Spectrum |
• FCC 47 CFR Part 15 Class C (2.4 GHz band US)
• FCC 47 CFR Part 15 Class E (5 GHz band US)
• RSS-247 (2.4 GHz and 5 GHz band Canada)
• ETSI EN 300 328 V2.1.1 (2.4 GHz band EU)
• ETSI EN 301 893 V2.1.1 (5 GHz band EU) |
Shock |
IEC 60068-2-27, 25 g |
Tripod Mounting |
UNC ¼″-20 on 2 sides |
Vibration |
• IEC 60068-2-6, 0.15 mm at 10–58 Hz and 2 g at 58–500 Hz, sinusoidal
• IEC 61373 Cat 1 (Railway) |
Corrosion |
• ISO 12944 C4 G or H
• EN60068-2-11 |
Files Storage & Data Transfer |
Video Storage |
Records up to 10 H.264 videos; storage as function of alarm — 5 sec. before alarm and 5 sec. after alarm. |
Imaging & Optical |
Available Lenses |
6°, 24°, 14°, 42° |
Detector Pitch |
24 µm |
Focal Plane Array (FPA) |
Uncooled microbolometer |
Focus |
One-shot contrast, motorized, and manual |
Image Frequency |
30 Hz |
Radiometric IR video streaming |
NA |
Readout |
Measurement results:
• Ethernet/IP (pull)
• Modbus TCP slave (pull)
• MQTT (push)
• Query over REST API (pull)
• Measurements and still image (radiometric JPEG, visual 640 × 480, visual 1280 × 960), read access only
• Web interface |
Spectral Range |
7.5–14 µm |
Streaming Resolution |
Video stream 0: 640 × 480 pixels Video stream 1: 1280 × 960 pixels |
Video Streaming |
RTSP protocol |
Overlay |
With / Without |
Visible Camera |
Optional accessory |
Measurement & Analysis |
External Optics & Windows Correction |
Based on input of optics/window transmission and temperature |
Measurement Corrections |
Global object parameters; local parameters per analyze function |
Measurement Presets |
Yes |
Reflected apparent temperature correction |
Based on input of reflected temperature |
Measurement Tools |
10 spotmeters, 10 boxes, 3 Deltas (difference any value/reference/external lock), 1 isotherm (above/below/interval), 1 iso-coverage, 1 reference temperature |
Automatic hot & cold detection |
Max./min. temperature value and position shown within box |
POWER |
External Power Operation |
24/48 V DC 8 W max |
Power Connection |
M12 12-pin A-coded, male (shared with Digital I/O) |
Power Consumption |
• 7.5 W at 24 V DC typical
• 7.8 W at 48 V DC typical
• 8.1 W at 48 V PoE typical |
External Voltage |
Allowed range 18–56 V DC |
Radiometry |
Image Source |
Video stream 0: visual / IR / MSX® (visual camera is an optional feature) Video stream 1: visual |
Storage Media |
Image Storage |
Records up to 100 FLIR radiometric JPEG; storage as function of: alarm, scheduling, or user interaction (camera web) |