Dwyer AQTIAP-AP2 Specifications |
AP2 THERMO-ANEMOMETER PROBE SPECIFICATIONS |
Service |
Dry, clean air. |
Temperature Limits |
Process: -20 to 200°F (-29 to 93°C); Ambient: -20 to 140°F (-29 to 60°C); Battery Charging: 32 to 113°F (0 to 45°C). |
Range |
Air Velocity: 0 to 6000 FPM (0 to 30 m/s); Volumetric Air: 999,999 in selected flow units; Temperature: -40 to 212°F (-40 to 100°C). |
Accuracy |
Air Velocity: ±3% FS within temperature range of 40 to 90°F (4 to 32°C); Temperature: ±0.5°F (±0.28°C). |
Response Time |
1 s. |
Probe Length |
8" (203 mm) insertion. |
Power Requirements |
3.7 V YT562447 Lithium ion battery, installed functional, user replaceable. |
Maximum Wireless Distance |
50' (15 m). |
Handle Enclosure |
Thermoplastic elastomer over polycarbonate. |
Supplied With |
Wrist strap. |
Weight |
11.2 oz (331.22 g). |
Agency Approvals |
CE (not while charging), FCC compliant. |
UHH2 SPECIFICATIONS |
Display |
4.3" QHD Gorilla glass, 960x540. |
Wireless Distance |
50' (15 m) or greater. |
Response Time |
1 s. |
Temperature Limits |
Operating: -4 to 140°F (-20 to 60°C); Storage: -40 to 176°F (-40 to 80°C). |
Humidity Limits |
5 to 95% RH. |
Power Requirements |
2000mAh lithium ion battery, installed functional, non-replaceable. |
Memory |
RAM 1G & ROM 4G. |
Operating System |
Android® 4.2.2. |
CPU |
MTK6589W Quad Core 1.2 GHz. |
Enclosure Rating |
IP68. |
Weight |
15.9 oz (450 g). |
Agency Approvals |
CE. |
RP2 THERMO-HYGROMETER PROBE SPECIFICATIONS |
Service |
Clean air. |
Temperature Limits |
Process: -4 to 140°F (-20 to 60°C); Ambient: 5 to 125°F (-15 to 51°C); Battery Charging: 32 to 113°F (0 to 45°C). |
Range |
RH: 0 to 100% (non-condensing); Temperature: -22 to 140°F (-30 to 60°C). |
Accuracy |
RH: ±2% @ 25°C (10 to 90% RH); ±4% (0 to 10, 90 to 100% RH); Temperature: ±0.54°F @ 77°F (±0.3°C @ 25°C). |
Response Time |
1.5 s. |
Probe Length |
8" (203 mm) insertion. |
Power Requirements |
3.7 V YT562447 Lithium ion battery, installed functional, user replaceable. |
Maximum Wireless Distance |
50' (15 m). |
Handle Enclosure |
Thermoplastic elastomer over polycarbonate. |
Supplied With |
Wrist strap. |
Weight |
11.2 oz (331.22 g). |
Agency Approvals |
CE (not while charging), RoHS, FCC compliant. |
VP2 VANE THERMO-ANEMOMETER PROBE SPECIFICATIONS |
Service |
Clean air. |
Temperature Limits |
Process: -4 to 140°F (-20 to 60°C); Ambient: 5 to 125°F (-15 to 51°C). |
Range |
Air Velocity: 50 to 5000 fpm (0.25 to 25 m/s); Temperature: -22 to 140°F (-30 to 60°C); Relative Humidity: 0 to 100% RH; Air Volume: 999,999 in selected units. |
Accuracy |
Air Velocity: 0.25 to 10 m/s: ±1.5% of reading ±20 fpm (±0.1 m/s); 10 to 20 m/s: 1.5% of reading ±40 fpm (±0.2 m/s); 20 to 25 m/s: ±1.5% of reading ±60 fpm (±0.3 m/s); Temperature: ±0.54°F @ 77°F (±0.3°C @ 25°C); Relative Humidity: ±2% @ 77°F (25°C) (10 to 90% RH); ±4% (0 to 10% and 90 to 100%). |
Response Time |
Air Velocity: 1 s; Temperature: 1.5 s; Relative Humidity: 1.5 s; Air Volume: 1 s. |
Probe Length |
8" (203 mm) insertion. |
Power Requirements |
3.7 V YT562447 Lithium ion battery, installed functional, user replaceable. |
Battery Charging Limits |
32 to 113°F (0 to 45°C). |
Maximum Wireless Distance |
50' (15 m). |
Vane Material |
Anondized aluminum. |
Handle Enclosure |
Thermoplastic elastomer over polycarbonate. |
Supplied With |
Wrist strap. |
Weight |
13.6 oz (402.20 g). |
Agency Approvals |
CE (not while charging), RoHS, FCC compliant. |
WDPM SPECIFICATIONS |
Service |
Non-corrosive dry gases. |
Wetted Materials |
Consult factory. |
Accuracy |
±0.5% FS span @ 25°C (includes non linearity, hysteresis, and non repeatability). |
Pressure Limits |
See model chart on catalog page. |
Engineering Units |
Pressure: in w.c., ft w.c., in Hg, psi, OzSl, mm w.c., cm w.c., mBar, Pa, hPa, and kPa; Velocity: fpm, mph, kn, m/s, m/h, k/h, and fps; Flow: cfm, gpm, gph, gpd, m3/s, m3/h, lps, lpm, and lph. |
Temperature Limits |
Compensated: 32 to 140°F (0 to 60°C); Process/Ambient: 14 to 140°F (-10 to 60°C). |
Thermal Effects |
±0.01% FS/°F (±0.02% FS/°C). |
Battery Charging Limits |
32 to 113°F (0 to 45°C). |
Power Requirements |
3.7 V YT562447 lithium ion battery, installed functionally, user replaceable. |
Wireless Distance |
At least 50' (15 m). |
Connections |
Two barbed connections for use with 1/8" (3.18 mm) or 3/16" (4.76 mm) ID tubing. |
Weight |
2.5 oz (70.87 g). |
Agency Approvals |
CE with CE approved charger, RoHS, FCC. |