Hops Meter With Carrying Case
The Delmhorst G-34 Hops Moisture Meter lets you easily check moisture levels in baled hops to assure high-quality harvest and safe storage. This meter has a moisture range of 7%-14% in hops. The large analog display and built-in battery check make this a favorite among hop growers worldwide. The G-34 meter cannot be used unless you purchase a Delmhorst Electrode, preferably the Delmhorst 30-E/C because it has 9" pins for greater penetration. Another good option for Hops is the 832 Long Pin Prod, which has 6 pins, each with 3 1/4" penetration.
The Delmhorst G-34 analog moisture meter is an essential tool for producing and storing baled hops. With a large analog display, built-in battery and calibration checks, and a rugged construction, this meter assures quality readings for post-drying inspection. The G-34 can be used with various electrodes in order to provide optimal moisture checking for baled hops.
Operation:
- To check battery voltage, push the switch down to the "OFF" position and hold. The batteries are OK if the needle moves beyond "11" (BATT OK"). The 9-volt batteries are located in the compartment on the case bottom.
- To check the meter calibration, push the switch to the "ON" position and hold. The meter is in calibration if it reads "13.5" on the dial. A tolerance of + 0.5 is acceptable. An optional external resistance standard (part no.MCS-44) is also available. Please contact Delmhorst or your local dealer to purchase.
- To turn the meter on, push the switch to the "ON" position and release. The meter remains on for 4 minutes. The auto-timer is reactivated for 4 minutes with each subsequent test with a reading of 8% or higher. If readings are below 8%, push the switch "ON" from time to time to prevent the meter from turning "OFF" automatically. The meter may be turned off manually by pushing the switch to the "OFF" position. When the meter is "ON", the needle rests below the 6% mark.
- To check the %MC of bales connect the 830-2 (10 in) or 830-5 (12 in) prod to the H-4 handle. Connect the H-4 to the input connector marked "electrode" on the meter. Force the probe into the bale and read the moisture content on the meter scale. An optional electrode (30-E/C) with two 9-1/2in insulated pins is also available.
Care of your Meter:
- Store your meter in a clean, dry place. The protective carrying case provided is an ideal storage place when the meter is not in use. If the meter has been left in a hot or cold environment overnight or for an extended period, allow the meter to acclimate to the temperature conditions in which it will be used for minimum 1-2 hours, or as long as possible.
- Clean the meter and electrodes with any biodegradable cleaner. Use the cleaner sparingly and on external parts only. Keep cleaner out of the sensor input connector.
- Remove the batteries if the meter will not to be used for one month or longer.