General Specifications |
Input voltage |
90 – 264 V (50/60 Hz) or DC 12 V |
Power consumption |
max. 70 VA |
Rechargeable battery (option) |
Lead-acid battery, 2 x 6 V / 6.5 Ah |
Battery life (option) |
Approx. 8 hours (self-sufficient operation) |
Display |
LCD colour display (320 x 240) |
Data interface |
USB 2.0 (type B plug)
BAUR Report Manager external USB interface (type A plug) |
Printer |
Matrix printer, 24 characters, 57 mm plain paper |
Ambient temperature (operational) |
-10°C to +55° |
Storage temperature |
-20°C to +60°C |
Humidity |
Non-condensing |
Dimensions (W x H x D) |
476 x 372 x 340 mm (closed)
476 x 635 x 420 mm (open) |
Weight |
Approx. 27 kg (without battery) Approx. 29 kg (with battery) |
Degree of protection |
IP 32 |
Safety and EMC |
CE-compliant in accordance with
Low Voltage Directive (2014/35/EC),
EMC Directive (2014/30/EC),
EN 60068-2-ff Environmental testing |
Software available in |
German, English, French, Spanish,
Portuguese, Italian, Russian, Czech,
Polish, Dutch, Chinese (CN),
Chinese (TW), Korean |
Insulating Oil Testing |
Output voltage |
DPA 60 C: 0 - 60 kVrms symmetrical
DPA 75 C: 0 - 75 kVrms symmetrical |
Voltage slew rate |
0.5 - 10 kV/s |
Switch-off time |
< 10 μs |
Voltage slew monitoring |
Real Breakdown Monitoring (RBM) |
Accuracy |
DPA 60 C: 0 – 60 kV ±1 kV
DPA 75 C: 0 – 75 kV ±1 kV |
Resolution |
0.1 kV |
Internal temperature recording of the oil sample |
0 - 99°C |
Temperature resolution |
1°C |
Test standards |
ASTM D1816:2012 1 mm,
ASTM D1816:2012 2 mm,
ASTM D1816/97,
ASTM D877/D877M:2013 PA,
ASTM D877/D877M:2013 PB,
BS EN 60156, CEI EN 60156,
CSSR RVHP:1985, IEC 60156:1995,
IRAM 2341:1972, JIS C2101:2010,
PN 77/E-04408, SEV EN 60156,
UNE EN 60156, NF EN 60156,
SABS EN 60156,
VDE 0370 Part 5:1996,
AS 1767.2.1 |
User-specific test sequences |
10 |