Programming
The IsoPure™ 96 can be programmed by using the color touch screen or the included mouse. After a protocol has been created, the instrument will calculate and display the estimated time for each step and the total time.
Typical step settings include:
Tip Comb Loading and Unloading: Automate tip comb loading and unloading by designating the respective plate positions in the script.
Mixing: Parameters for mixing time, speed, and amplitude can be programmed for each step of the protocol.
Temperature Control: Heating the sample mixture during the lysis and elution steps significantly improves the purity and yield of the target molecule. The IsoPure™ 96 is equipped with 2 heating blocks on the plate carousel that can be precisely set to temperatures from ambient + 5°C to 120°C. IsoPure™ deep well plates fit into the heating blocks for efficient heat transfer. A cooling fan inside the processing chamber maintains the ambient operating temperature to ensure consistent results.
Bead Collection and Release: The complete collection of magnetic beads from sample mixtures is crucial for maximizing the yield of purified analytes. The IsoPure™ software offers full control of magnetic parameters to ensure cleaner extractions. General settings allow control of magnet speed, position, repeat collection cycles, etc, to guarantee consistent yields.