Simpson 228 AC/DC Leakage Current Tester

Click image to enlarge
Simpson 228
|
|
Features:
- True RMS Readings
- Reads in Measurement Indication Units (MIU) up to .005 MIU
- Output Allows Measurement of Peak Current in Non-Sinusoidal Wave Forms
- Let-Go, Reaction, and Burn Hazard Response Networks
- Burn Hazard Network
- Incorporates 3-Special Networks Required by Guidelines
- Detects Dangerous AC & DC Leakage Currents
|
Description
| Test for dangerous leakage current with the Simpson Model 228, designed to meet the latest UL, IEC, and ANSI specifications. The 228 detects both AC and DC leakage current and simulates human response at three different levels -- reaction, let-go, and burn hazard. Each Model 228 comes with test leads with alligator clips, manual, and batteries. |
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 228 |
228 .... IEC AC LEAKAGE TESTER (Item# 40027) |
$613.70 |
| $583.02 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| 00125 |
PROBE SET, THREADED W/CLIPS, BANAN |
$18.76 |
| $17.82 |  |
|
| 00834 |
CASE, POLYESTER PADDED, BLACK, 260 |
$59.13 |
| $56.18 |  |
|
Specifications
| Equivalent measuring instrument load |
| Reaction Response Network |
1 MΩ - 10pF |
| Let-Go Response Network |
1 MΩ - 62pF |
| Unweighted Burn Hazard Response Network |
1 MΩ - 1pF |
| Ranges |
| Reaction Response Network |
.3, 1, 3, 10 M.I.U. |
| Let-Go Response Network |
.3, 1, 3, 10 M.I.U. |
| Unweighted Burn Hazard Response Network |
0-100 mA RMS |
| Meter accuracy |
±2% F.S. @ 60 Hz |
| Meter frequency response |
| DC to 1 Hz |
Pointer shall track within 5% of peak |
| 15 Hz to 200 KHz |
%plusmn;2% of F.S. |
| 200 KHz to 1 MHz |
±2% F.S., ±2% F.S., ±5%F.S. |
| Output |
| Sensitivity |
Full scale meter indication equals 1Ω RMS (Measured with a 1 MV, 12 pF load) |
| Accuracy |
±2% of reading @ 60 Hz |
| Output frequency response |
| Type |
0.52" (13.2mm) Diameter Electret Condenser, Omnidirectional 70° angle of incidence. Maximum sound pressure level 148dB |
| Display |
| DC to 1Hz |
±2% of reading |
| 50 Hz to 100 KHz |
±2.5% of reading, +2%/-3% of reading, ±2% of reading |
| 100 kHz to 1 MHz |
±5% of reading |
| Voltmeter |
| Range |
0-300 V (AC or DC) |
| Accuracy: DC to 1 Hz |
Pointer shall track within 5% of peak |
| Accuracy: 20 Hz to 1KHz |
±3% F.S. @ 60 Hz |
| Frequency response |
DC to 1 KHz |
| General |
| Power requirements |
(2) 9V (NEDA 1604A) batteries |
| Physical |
7" x 5.25" x 3", 2-1/2 lbs (1.4kg), ABS plastic, fused input |
| Environmental |
27°C ±2°C, 70% noncondensing relative humidity. Outside this range derate by 10% per °C |
| Maximum operating range |
0° to 40° C |
|