Meriam RTD Measurement & Simulation Module

Click image to enlarge
Meriam RIO4000-11-1
|
|
Features:
- Measure or simulate temperature for 22 different RTD types
- Measure or simulate ohms
|
Description
RTD Measurement & Simulation Module - For Use with MFT 4000 Series Calibrator
The RIO4000 is the RTD measure and simulate module for the Meriam MFT 4000 Series Multi-Function Tester. The module accurately measures or simulates the resistance of 22 different RTD sensor types with nominal resistance to 4000 ohms. 2-, 3-, and 4-wire RTD Measurement for various RTD types; platinum, copper, nickel or iron. Various alpha coefficients are supported per Table 1. The module can simulate temperature to all conventional or smart temperature transmitters and signal conditioners.
The MFT converts measured resistance to a corresponding temperature value and displays the value in the user selected temperature engineering units (°F, °C, °R or °K). Ohms can also be displayed for direct reading or troubleshooting. Pt 100 RTDs probes are available from Meriam for use with the module. The Meriam RTD probes are 4-wire, DIN Class A probes, which come complete with handle, 5’ coiled cable, and connector. Accessory connectors are available to connect to any RTD or temperature transmitter for measurement or simulation applications. The RTD Simulation mode can be setup by entering a specific temperature or a range of temperatures that can be set up for stepping and ramping functions.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| RIO4000-11-1 |
RTD Measurement & Simulation Module |
| Call |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| A37063 |
Starter Kit for RIO4000 module RTD meas. & sim. - Universal; A900028-90500 Measure cable; module connector by four (4) alligator clips; A900028-90200 RTD mating connector (for use with customer's RTD); A900028-90502 Simulate Cable; module connector by two (2) aligator clips |
| $125.00 |  |
|
| A900028-90200 |
Mating connector for RTD temperature modules facilitates use of customer's RTD |
| $20.00 |  |
|
| A900028-90201 |
Pt 100 RTD, DIN Class "A", 0.00385 alpha, 4 lead, 1/4" O.D. 300 Series SS probe, 10" length, 3.5" handle and 5 ft coiled cable with connector for RIO4000 temperature module |
| $232.00 |  |
|
| A900028-90500 |
Measure Cable for RTDs temp module, 5 ft cable, 4 lead wires, alligator clip terminations |
| $105.00 |  |
|
| A900028-90502 |
Simulate Cable for RIO4000 Module temp module, 3 ft cable, 2 lead wires, alligator clips |
| $34.00 |  |
|
| B34686 |
961P pressure pump with integral variator, bleed valve, 1/8" FNPT connection, 0 - 145 PSIG |
| $376.00 |  |
|
| B34700 |
961V vacuum pump with integral variator, bleed valve, 1/8" FNPT connection, -650 mm Hg |
| $316.00 |  |
|
| A34386 |
Connector Kit: (not for use with GI2000) kit includes: 1/8" tubing, "T" unions (2), 1/8" MNPT adapters (6), 3/16" Barb to 1/8" FNPT (1) & 1/4" MNPT to 1/8" FNPT reducer (2) |
| $84.00 |  |
|
| A34102 |
1/8" O.D. tubing to 1/8" MNPT connector (not for use above 250 PSIG) |
| $8.00 |  |
|
| A34112 |
1/8" O.D. tubing union tee (not for use above 250 PSIG) |
| $11.00 |  |
|
| A34099-1 |
1/4" MNPT X 1/8" FNPT reducer bushing, brass |
| $7.00 |  |
|
| A34103-2 |
10 ft of 1/8" O.D. nylon tubing (not for use above 250 PSIG) |
| $15.00 |  |
|
| A36856 |
Please call or email us for pricing and availability. |
| A900020-90502 |
High Pressure Connection Kit (not for use above 9,000 PSIG) includes: 3' HP Hose with 1pc 1/8" MNPT & 1pc 1/4" MNPT x HP Quick Connect fitting |
| $113.00 |  |
|
| A900020-90503 |
High Pressure Connection Kit (not for use above 9,000 PSIG) includes: 1' HP Hose with 1pc 1/8" MNPT and 1pc 1/4" MNPT x HP Quick Connect fitting |
| $100.00 |  |
|
| A900020-90501 |
5 pcs 1/4" MNPT x HP Quick Connect Fitting (not for use above 9,000 PSIG) |
| $144.00 |  |
|
| A900436-00001 |
1 pcs 1/8" MNPT HP Quick Connect fitting (not for use above 9,000 PSIG) |
| $28.00 |  |
|
Specifications
| 0-400 Ω Measurement Accuracy |
±0.01% of R + 0.075 Ω |
| 0-400 Ω Simulation Accuracy |
See Table 2 |
| 400-4000 Ω Measurement Accuracy |
±0.01% of R + 0.237Ω |
| 400-4000 Ω Simulation Accuracy |
See Table 2 |
| Resistance Range |
0 - 4000 Ω |
| Input impedance |
> 10 megohms |
| Mating Connector |
Switchcraft #TA6FL Pulsed excitation current support |
| Units |
°F, °C, °R, °K, ohms |
| Temperature |
Effect: = 0.001 ohm/ oC Storage: -40°F to 140°F (-40°C to 60°C) Operating: 14°F to 122°F (-10°C to 50°C) |
| Resolution |
±0.1° or ±1° for RTDs |
| Weight |
3 ozs. |
| Certifications Available |
CE Mark (standard) NIST traceability certificate (standard) Intrinsically Safe, MET Laboratories per CSA C22.2 & UL Class I Division I (standard) |
Table 1: Supported RTD Materials / Nominal Resistance |
Material |
R0 |
ALPHA |
Rs(ID) |
Temp Range (°C) |
| Platinum |
25.5 |
0.00392 |
17.8 |
-200 to +630° C |
| Platinum |
98.129 |
0.00392 |
21.5 |
-200 to +600° C |
| Platinum |
100 |
0.00385 |
100 |
-200 to +850° C |
| Platinum |
100 |
0.00392 |
46.4 |
-200 to +630° C |
| Platinum |
100 |
0.00391 |
56.2 |
-200 to +630° C |
| Platinum |
100 |
0.00393 |
68.1 |
-200 to +850° C |
| Platinum |
100 |
0.00389 |
82.5 |
-200 to +630° C |
| Platinum |
200 |
0.00385 |
178 |
-200 to +630° C |
| Platinum |
200 |
0.00392 |
215 |
-200 to +630° C |
| Platinum |
470 |
0.00392 |
261 |
-200 to +630° C |
| Platinum |
500 |
0.00392 |
316 |
-200 to +630° C |
| Platinum |
500 |
0.00391 |
383 |
-200 to +630° C |
| Platinum |
500 |
0.00385 |
464 |
-200 to +630° C |
| Platinum |
1000 |
0.00385 |
1000 |
-200 to +630° C |
| Platinum |
1000 |
0.00375 |
1210 |
-50 to +500° C |
| Copper |
9.035 |
0.00427 |
12.1 |
-100 to +260° C |
| Copper |
100 |
0.00427 |
121 |
-100 to +260° C |
| Nickel |
100 |
0.00618 |
31.6 |
-60 to +160° C |
| Nickel |
120 |
0.00672 |
147 |
-80 to +260° C |
| Iron |
604 |
0.00518 |
562 |
-100 to +200° C |
| Iron |
908.4 |
0.00527 |
681 |
-100 to +200° C |
| Iron |
1816.81 |
0.00527 |
1470 |
-100 to +200° C |
The table above lists the RTDs supported by the RIO4000 module and MFT 4000 Series.
Materials, types and alpha coefficients are based on popular RTDs used in the field.
Table 2: RIO4000 Accuracy Table |
| |
Operating Temperature 18 < T < 28° C* |
RTD Type |
(á) |
T Range (°C) |
Measure (2, 3, 4 wire) |
Simulate (2 wire**) |
Pt 100 |
385 |
-200 to 0 |
0.3 |
0.3 |
0 to 400 |
0.4 |
0.4 |
400 to 850 |
0.5 |
0.6 |
Pt 100 |
389, 391 |
-200 to 0 |
0.4 |
0.4 |
0 to 360 |
0.4 |
0.6 |
Pt 100 |
393 |
-200 to 0 |
0.5 |
0.4 |
0 to 400 |
0.4 |
0.5 |
400 to 850 |
.6 |
0.6 |
Pt 200 |
385, 392 |
-200 to 0 |
0.2 |
0.4 |
0 to 400 |
0.6 |
0.5 |
400 to 630 |
0.8 |
0.6 |
Pt 500 |
385, 392 |
-200 to 0 |
0.2 |
0.4 |
0 to 400 |
0.3 |
0.5 |
400 to 630 |
0.4 |
0.6 |
Pt 500 |
391 |
-200 to 0 |
0.2 |
0.4 |
0 to 400 |
0.4 |
0.5 |
400 to 630 |
0.5 |
0.6 |
Pt 1000 |
385 |
-200 to 0 |
0.2 |
0.4 |
0 to 400 |
0.3 |
0.5 |
400 to 630 |
0.4 |
0.6 |
Pt 1000 |
375 |
-50 to 0 |
0.3 |
0.4 |
0 to 500 |
0.3 |
0.5 |
Cu 10 |
4274 |
-100 to 0 |
2 |
2 |
0 to 260 |
2 |
2 |
Cu 100 |
4274 |
-100 to 0 |
1 |
1 |
0 to 260 |
1 |
1 |
Ni 100 |
618 |
-60 to 160 |
0.2 |
0.4 |
Ni 120 |
672 |
-80 to 260 |
0.2 |
0.4 |
Fe 604 |
518 |
-100 to 200 |
0.3 |
0.5 |
Fe 908 |
527 |
-100 to 200 |
0.3 |
0.5 |
Fe 1816 |
527 |
-100 to 200 |
0.2 |
0.5 |
Ambient temperature limits: -10º C to +50º C
Accuracy statements are for ambient temperatures of 18º C to 28º C.
Apply the Temperature Coefficient for ambient temperatures below 18º C and above 28º C
*Temperature Coefficient = 0.02 ºC / ºC
** For 2-Wire Simulation, Manufacturer’s 2 Wire Simulation Cable is Required.
back to top
|