Series BG |
Series CG |
Series EG |
Series MG |
Series 5 |
|
|
|
|
|
|
Accuracy (% of full scale) |
±0.2% |
±0.3% |
±0.5% |
0.1% of full scale 1 digit |
Sampling rate (reading/s) |
65 |
30 |
|
Display rate (updates/s) |
2.5 / 65 in peak mode |
2.5 / 30 in peak mode |
|
Power |
AC or rechargeable battery (both included). Includes Intelligent Power Management System (IPM). IPM warns the user via a "LO BAT" indicator and then will power off the gauge. |
AC or non-rechargeable battery, includes IPM |
|
Battery life |
8-10 hours of continuous use |
30 hours of continuous use |
Backlight on: up to 7 hours of continuous use
Backlight off: up to 24 hours of continuous use |
Peak capture |
Yes, in tension and compression (cw and ccw for torque gauges) |
|
Units of measurement |
Force gauges: lb(oz), kg(g), N
Torque gauges: lbin(ozin), kgmm, Ncm for torque gauges |
|
Outputs |
Standard |
Optional |
N/A |
Advanced |
RS-232 |
Fully configurable up to 9600 baud. Includes Gauge Control Language (GCL). |
Fully configurable up to 9600 baud. |
N/A |
Configurable up to 115,200 baud. Includes Gauge Control Language 2 for full computer control. |
Mitutoyo (Digimatic) |
Serial BCD suitable for all Mitutoyo SPC-compatible devices |
N/A |
Serial BCD suitable for all Mitutoyo SPC-compatible devices. |
Analog |
±1 VCD, ±0.25% of full scale at capacity. Positive for compression, negative for tension. |
N/A |
1 VCD, 0.25% of full scale at capacity, |
General purpose I/O |
Three open collector outputs and one input (utilizing Mitutoyo lines) |
N/A |
Three open drain outputs, one input. |
Set point |
Three open collector lines (utilizing Mitutoyo lines) |
N/A |
Three open drain lines. |
Configurable Settings |
Analog filters, digital filters, outputs selection, automatic output (RS-232), automatic shutoff, initial (default), averaging mode, calibration |
Outputs selection, automatic shutoff, initial (default), calibration |
N/A |
Digital filters, outputs, automatic output (via USB/RS-232), automatic shutoff, default settings, averaging mode, external trigger, passwords, key tones, audio alarms, backlight, calibration |
Load cell deflection |
0.010 in [0.25 mm] at full capacity |
0.010 [0.25], except 0.050 [0.13] |
Safe overload (% of full scale) |
150% (display shows "----" at 110% and above) |
Weight (gauge only) |
BG012 - BG200: 0.95 lb [0.4 kg] BG500: 1.12 lb [0.5 kg] |
All capacities: 1.8 lb [0.8 kg] |
EG012 - EG200: 0.95 lb [0.4 kg] EG500: 1.12 lb [0.5 kg] |
MG025 - MG200: 0.7 lb [0.3 kg] MG500: 0.87 lb [0.4 kg] |
M5-012 - M5-100: M5-012 - M5-100: 1.0 lb [0.45 kg]
M5-200 - M5-500: 1.2 lb [0.54 kg] |
Included accessories |
Chisel, cone, V-groove, hook, flat, extension rod |
Hook, flat |
Carrying case with chisel, cone, V-groove, hook, flat, extension rod, universal voltage AC adapter, battery, quick-start guide, USB cable, resource CD (USB driver, MESURgauge DEMO software, and users guide), and NIST-traceable certificate of calibration with data. |
Environmental requirements |
40°F - 100°F [5°C - 45°C], <96% humidity (non-condensating) |
|
Thermal effects |
Zero: 0.03% of full scale/°C, Span: 0.01% of full scale/°C |
|
Housing construction |