Mark-10 Force/Torque Comparison

Description

The following table summarizes the technical characteristics of Mark-10 force gauges in an easy-to-read comparison format. For a comparison of test stands, click here. Note: The BGI force/torque gauge shares its features and specifications with the BG (except for load cell-related information, weight, and included accessories). The MGT torque gauge also shares most of its characteristics with the MG force gauge, with the same exceptions. Force and Torque Gauges Comparison Chart (printable .pdf version)

  Series BG Series CG Series EG Series MG Series 5
 
Accuracy (% of full scale) ±0.2% ±0.3% ±0.5% 0.1% of full scale 1 digit
Sampling rate (reading/s) 65 30  
Display rate (updates/s) 2.5 / 65 in peak mode 2.5 / 30 in peak mode  
Power AC or rechargeable battery (both included). Includes Intelligent Power Management System (IPM). IPM warns the user via a "LO BAT" indicator and then will power off the gauge. AC or non-rechargeable battery, includes IPM  
Battery life 8-10 hours of continuous use 30 hours of continuous use Backlight on: up to 7 hours of continuous use
Backlight off: up to 24 hours of continuous use
Peak capture Yes, in tension and compression (cw and ccw for torque gauges)  
Units of measurement Force gauges: lb(oz), kg(g), N
Torque gauges: lbin(ozin), kgmm, Ncm for torque gauges
 
Outputs Standard Optional N/A Advanced
RS-232 Fully configurable up to 9600 baud. Includes Gauge Control Language (GCL). Fully configurable up to 9600 baud. N/A Configurable up to 115,200 baud. Includes Gauge Control Language 2 for full computer control.
Mitutoyo (Digimatic) Serial BCD suitable for all Mitutoyo SPC-compatible devices N/A Serial BCD suitable for all Mitutoyo SPC-compatible devices.
Analog ±1 VCD, ±0.25% of full scale at capacity. Positive for compression, negative for tension. N/A 1 VCD, 0.25% of full scale at capacity,
General purpose I/O Three open collector outputs and one input (utilizing Mitutoyo lines) N/A Three open drain outputs, one input.
Set point Three open collector lines (utilizing Mitutoyo lines) N/A Three open drain lines.
Configurable Settings Analog filters, digital filters, outputs selection, automatic output (RS-232), automatic shutoff, initial (default), averaging mode, calibration Outputs selection, automatic shutoff, initial (default), calibration N/A Digital filters, outputs, automatic output (via USB/RS-232), automatic shutoff, default settings, averaging mode, external trigger, passwords, key tones, audio alarms, backlight, calibration
Load cell deflection 0.010 in [0.25 mm] at full capacity 0.010 [0.25], except 0.050 [0.13]
Safe overload (% of full scale) 150% (display shows "----" at 110% and above)
Weight (gauge only) BG012 - BG200: 0.95 lb [0.4 kg] BG500: 1.12 lb [0.5 kg] All capacities: 1.8 lb [0.8 kg] EG012 - EG200: 0.95 lb [0.4 kg] EG500: 1.12 lb [0.5 kg] MG025 - MG200: 0.7 lb [0.3 kg] MG500: 0.87 lb [0.4 kg] M5-012 - M5-100: M5-012 - M5-100: 1.0 lb [0.45 kg]
M5-200 - M5-500: 1.2 lb [0.54 kg]
Included accessories Chisel, cone, V-groove, hook, flat, extension rod Hook, flat Carrying case with chisel, cone, V-groove, hook, flat, extension rod, universal voltage AC adapter, battery, quick-start guide, USB cable, resource CD (USB driver, MESURgauge DEMO software, and users guide), and NIST-traceable certificate of calibration with data.
Environmental requirements 40°F - 100°F [5°C - 45°C], <96% humidity (non-condensating)  
Thermal effects Zero: 0.03% of full scale/°C, Span: 0.01% of full scale/°C  
Housing construction 6061 aluminum, reversible halves for handheld operation or test stand mounting  