Graphtec GL900 GL Series midi LOGGER

Click image to enlarge
Graphtec GL900
|
|
Features:
- Multifunction input on eight isolated channels
- High-speed simultaneous sampling on eight channels, 16-bit resolution
- Equipped with a large-format 5.7-inch color LCD for easy-to-read waveform display
- Data can also be saved to PC-friendly USB memory sticks
- Built-in Ethernet 100Tx and USB 2.0 interfaces
- Accepts voltages, thermocouples and humidity
- 8 channels
- Voltage input levels: 20mV to 500V F.S.
- Thermocouple input types: B, R, S, T, N, K, W, E, J
- Continuous battery charge during operation
- 256MB internal flash memory
- Includes 4 pulse logic inputs, 1 trigger input
- Includes 4 x alarm outputs
- Compact data logger footprint (232mm x 150mm x 80mm)
- Includes Windows software for setup, control, data transfer, data conversion to ASCII and Excel formats, and on-screen data review with cursors
|
Description
|
In compliance with various test requirements, this data logger is capable of performing high-speed simultaneous voltage and temperature measurements
Easy-to-use upright high-speed isolated 8-channel multifunction logger
An easy-to-use upright device enabling isolated 8-channel multifunction input, the GL900 is capable of performing high-speed simultaneous measurements of voltage, temperature, and various other phenomena.
High-voltage measurement capability
The wide 500 V range enables 100 to 240 VAC power supply voltage waveform measurements. Using logic input and a clamp meter simultaneously allows measurement of a device’s power supply voltage and current concurrently with sequential control of various points.
Built-in, large-format 5.7-inch color LCD for easy-to-read waveforms
The bright, easy-to-read large-format 5.7-inch color TFT LCD provides vivid, easy-to-read waveform displays. Cursor keys enable fast, easy control and setup. The waveform display can be scrolled at high-speed – 10 ms/DIV.
Free Running display for waveform-checking without the need for data capture
The Free Running display lets users check input signal waveforms even before measurements begin. Since waveforms are displayed on each setup screen, users can make settings while viewing the waveforms.
Data can be captured to PC-friendly USB memory sticks
Long-term data can be captured directly to built-in 256-MB flash memory or to an external USB memory stick at sampling intervals of from 1 ms to 1 min. For high-speed sampling at intervals faster than 1 ms, up to one million data points can be captured to internal RAM.
Easy PC measurement via USB; remote monitoring via Ethernet web server and FTP functions
The USB and Ethernet connections enable transfer of captured data to your PC and setup and control of the GL900 from a PC, even without the PC software provided standard with the GL900.
Can be used as an X-Y recorder
The GL900 reproduces analog X-Y recorder movements and provides the illusion of pen up/pen down movements. It can be operated like an analog X-Y recorder and can also be used as a 4-pen X-Y recorder. The digital data format facilitates post-measurement confirmation of data values and report creation.
High-precision temperature measurement even during high-speed sampling
Lets users perform high-precision temperature measurements even during high-speed sampling – ideal for performing combined voltage and temperature measurements.
Dedicated software for real-time data capture
Three measurement screens are provided to allow selection of the screen that best suits measurement needs. The Replay screen provides a Zoom screen feature to enable enlarged display of specific sections of long-term measurement data.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| GL900 |
High-speed isolated 8-channel multifunction logger |
$3995.00 |
| $3915.10 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| RIC-141 |
Please call or email us for pricing and availability. |
| RIC-112 |
BNC-BNC cable, 6 ft |
$35.00 |
| $34.30 |  |
|
| RIC-113 |
BNC-Banana cable, 6 ft |
$36.00 |
| $35.28 |  |
|
| RIC-114 |
BNC-Alligator Clips cable, 6 ft |
$27.00 |
| $26.46 |  |
|
| B-517 |
Battery pack (installs within the mainframe, powers the system for approximately 6 hours with two batteries) |
$120.00 |
| $117.60 |  |
|
| B-513 |
Cable: 4 x Logic inputs & 4 x alarm outputs, 2 meters |
$80.00 |
| $78.40 |  |
|
| B-514 |
DC power cable, 2 meters, bare ends, for direct connection to DC power source |
$50.00 |
| $49.00 |  |
|
| B-536 |
Midi Logger carrying case for GL200 / GL800 |
$400.00 |
| $392.00 |  |
|
| B-530 |
Humidity sensor, 0 to 100% RH |
$400.00 |
| $392.00 |  |
|
| RIC-410 |
Please call or email us for pricing and availability. |
| RIC-420 |
Please call or email us for pricing and availability. |
| RIC-430 |
Please call or email us for pricing and availability. |
| CM-211 |
Please call or email us for pricing and availability. |
| CM-114 |
Please call or email us for pricing and availability. |
| CM-113 |
Please call or email us for pricing and availability. |
Specifications
| No. of analog input ch. |
8 ch |
| External input/output |
Trigger input (1 channel), Logic input (4 channels) or Pulse input (4 channels), Alarm output (4 channels) |
| Sampling interval |
10µs to 1 min |
| TIME/DIV |
10 ms/DIV to 24 hour/DIV |
| Timer functions |
Date and time, daily cycle, hourly cycle |
| Trigger Function |
Type |
Start (Data capture starts when a trigger is generated) Stop (Data capture stops when a trigger is generated) |
| Condition |
Start: Off, Input signal level (analog, logic/pulse), External*1 Stop: Off, Input signal level (analog, logic/pulse), External *1, Scheduled time |
| Combination |
Input signal level: Level OR, Level AND, Edge OR, Edge AND |
| Mode |
H (Rising), L (Falling), Window In*2, Window Out*2 |
| Alarm setting functions |
Rising, Falling, Window In*2, Window Out*2 |
| Alarm output *1 |
Number of channels: 4, Open collector output (5V, 10 kO pull-up resistance) |
| Pulse Input |
RPM mode |
5 to 20 M RPM/F.S. (in steps of 1, 2, or 5) |
| Count mode |
5 to 20 M C/F.S. (in steps of 1, 2, or 5) |
| Inst. Mode |
5 to 20 M C/F.S. (in steps of 1, 2, or 5) |
| Calculation functions |
Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously) |
| Other functions |
Search function, annotation input function |
| PC inteface |
Ethernet (10BASE-T/100BASE-TX), USB (High Speed supported) provided as standard |
| Ethernet functions |
Web server function, FTP server function, NTP client function |
| USB function |
USB drive mode (File transfer and deletion from internal GL900 memory) |
| Memory device |
Internal |
One million data points / Internal flash memory:Approx. 256 MB |
| External |
USB memory slot (High speed supported) *5 |
| Display screens |
Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y |
| Display unit |
5.7-inch TFT color LCD |
| Operating environment |
0 to 40°C, 5 to 85% R.H. (15 to 35? when using batteries) |
| Withstand voltage |
Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute |
| Power supply |
AC adapter |
100 to 240 VAC, 50 to 60 Hz |
| DC input |
8.5 to 24 VD |
| Battery pack *6 |
Option |
| Power consumption |
30 VA (when operating with AC power) |
| External dimensions |
232 x 150 x 80 mm (W x H x D), approx. |
| Weight (approx.) |
1.1 kg (excluding AC adapter and battery) |
| Vibration-tested conditions |
Equivalent to automobile parts Type 1 Category A classification |
GL900 Input Terminal Unit Specifications
| Input terminal type |
Voltage |
BNC connector |
| Temperature |
M3 screw type terminal board *7 |
| Method |
All channels isolated Imbalanced input Simultaneous sampling of all channels |
| Measurment Range |
Voltage |
20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100, 200, 500 V F.S., 1-5 V F.S. |
| Temperature |
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26) |
| Humidity |
0 to 100% (voltage 0 V to 1 V scaling conversion) with B-530 (option) |
| Input filter |
Off, Line, 5 Hz, 50 Hz, 500 Hz |
Measurement Accuracy (23°C ± 5°C) When 30minutes or more have elapsed after power was switched on Sampling 1s/20ch Filter ON(10) GND connected |
Voltage |
±0.25% of F.S. |
| Temperature |
Measurement range |
Accuracy |
| Thermocouple |
R/S |
0 ≤ TS ≤100 °C |
±7.0°C |
| |
100 < TS ≤300 °C |
± 5.0 °C |
| |
R : 300 < TS ≤1600 °C |
± (0.05 % of rdg + 3.0 °C) |
| |
S : 300 < TS ≤1760 °C |
± (0.05 % of rdg + 3.0 °C) |
| B |
400 ≤ TS ≤ 600 °C |
±5.5°C |
| |
600 < TS ≤ 1820 °C |
±(0.05% of rdg +3.0°C) |
| K |
-200 ≤ TS ≤-100 °C |
±(0.05% of rdg +3.0°C) |
| |
-100 < TS ≤1370 °C |
±(0.05% of rdg +2.0°C) |
| E |
-200 ≤ TS ≤-100 °C |
±(0.05% of rdg +3.0°C) |
| |
-100 < TS ≤800 °C |
± (0.05 % of rdg + 2.0 °C) |
| T |
-200 ≤ TS ≤ -100 °C |
± (0.1 % of rdg + 2.5 °C) |
| |
-100 < TS ≤400 °C |
± (0.1 % of rdg + 1.5 °C) |
| J |
-200 ≤ TS ≤ -100 °C |
± 3.7 °C |
| |
-100 ≤ TS ≤100 °C |
± 2.7 °C |
| |
100 < TS ≤1100 °C |
± (0.05 % of rdg + 2.0 °C) |
| N |
0 ≤ TS ≤1300 °C |
± (0.1 % of rdg + 2.0 °C) |
| W |
0 ≤ TS ≤2315 °C |
± (0.1 % of rdg + 2.5 °C) |
| * If the reference junction compensation is internal, add ± 1.0 °C to each of the above values. |
| A/D converter |
16 bit (out of which 14 are internally acknowledged) |
| Maximum permissible input voltage |
Between input channel + and - terminals |
20 mv to 1 V : ±30VDC 2 V to 500 V : ±500VDC |
| Between input channel terminals |
60 Vp-p |
| Between input channel terminal and GND terminal |
60 Vp-p |
| Withstand voltage |
Between input channel terminal and GND terminal |
1 minute at 1000 Vp-p |
| Between input channel terminals |
1 minute at 1000 Vp-p |
| *1 Logic alarm cable (B-513) is required. |
| *2 Cannot be set for logic input |
| *3 Maximum input frequency: 50 kHz, maximum number of counts: 15 M C |
| *4 In real time or when Between Cursors has been specified (during Replay) |
| *5 1 file = 2 Gbytes (depends on the USB memory stick used) *6 Please install two battery packs. |
| *7 Connections are made to both the BNC terminal and M3 screw terminal for the same channel. |
| *8 Thermocouple diameters T:0.32f, others:0.65f |
| *9 Operating temperature range: -25 to +80°C |
|