Features:
- The GL200 has a 3.5" color LCD display
- Built-in Ethernet and USB 2.0 interfaces
- Accepts voltage, temperature, humidity, pulse and logic inputs
- 10ch terminal stand with M3 screws
- 100 mS/s sampling rate max (10-ch. config.), or as slow as 1 S/hr
- Built-in 3.5 mb flash memory
- High input isolation is built-in! Enhanced safety and ease of connection.
- Pulse input: 1ch/ Logic input:1ch
- Ultra small A5 size (194 x 122 x 41 mm)
|
Description
|
10 isolated channels, multifunction input.
The isolated input system ensures that none of the channels is affected by different signals input to the other channels and eliminates the need to pay special attention to wiring. The multifunction GL200 device accepts input of voltage, temperature, humidity, pulse and logic signals. This allows combined measurements, even for disparate phenomena such as temperature/humidity and voltage.
Accepts USB memory devices of up to 1GB capacity and allows easy connection to PCs
Two ports are provided, one for the USB memory device and one for linking to a PC. Commercially-available USB memory devices may be used for direct capture of data, with a 512-MB USB memory device, users can capture data for up to approximately 269 days *1. In addition, connecting the GL200 to a PC allows measurements to be made from the PC. *1 10 channels, 1-s sampling rate
Vivid 3.5-inch TFT color LCD
The bright, easy-to-read 3.5-inch TFT color LCD monitor makes it even easier to review measurement parameter settings, measured waveforms, and digital values.
The product is bundled with PC application software that enables measurements with the GL200 from a PC. In addition to waveform monitoring, this software provides other useful functions such as direct writing of data to an Excel file and report creation. The software also provides a Help display function for enhanced usability.
Bundled PC Software
The unit is bundled with PC application software that enables measurements with the GL200 from a PC. In addition ro waveform monitoring, this software provides other useful functions such as direct writing of data to an Excel file and report creation. The software also provides a Help display function for enhanced usability.
Built-in flash memory for reliable measurements
The flash memory used for the built-in memory device ensures that data is retained even if the power supply is interrupted. This means momentary disruptions in the AC power supply or accidentally switching off power will not result in data loss.
| Data Capture Times (at the time of 10 ch measurement, pule & logic off) |
| Capture interval (Sampling intervals) |
100ms |
200ms |
500ms |
1s |
10s |
| 3.5MB-built in flash memory |
Approx.3.9hrs |
Approx.7.8hrs |
Approx.19hrs |
Approx.1.6days |
Approx.16days |
| 256MB USB memory device |
Approx.13days |
Approx.26days |
Approx.65days |
Approx.130days |
Approx.1300days |
6 hours of continuous battery operations
A convenient optional battery pack is available for use in areas where AC poser is unavailable or subject to interruption. The battery pack enables 6 hours* of continuous measurement and provides a remaining battery capacity display function, * Varies with measurement conditions.
M3-screw terminal block
The M3-screw terminal block used for the input section provides no-fuss connections and easy, secure wiring. Both round and open-ended terminals can be connected.
User friendly design
The GL200 incorporates feature such as the ability to set measurement parameters for each item separately and navigation assistance to provide superior operability. The device case features stylish non-slip rubber.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| GL200 |
GL200 midi LOGGER Mainframe - Data logger with 3.5" color display, with 10 channels, 3.5 MB on-board memory, USB PC interface, 16-bit resolution, high isolation per channel (125 VAC RMS and 350 VDC channel-to-chassis) |
$1250.00 |
| $1225.00 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| B-514 |
DC power cable, 2 meters, bare ends, for direct connection to DC power source |
$50.00 |
| $49.00 |  |
|
| B-517 |
Battery pack (installs within the mainframe, powers the system for approximately 6 hours with two batteries) |
$120.00 |
| $117.60 |  |
|
| B-530 |
Humidity sensor, 0 to 100% RH |
$400.00 |
| $392.00 |  |
|
Specifications
| Number of analog inputs |
10 channels |
| Input Method |
Photo MOS relay scanning system; all channels isolated |
| Measurment Ranges |
Voltage |
20, 50, 100 500mV; 1, 2, 5, 10, 20, 50V ; 1-5 v F.S. |
| Temperature |
Thermocouples: K, J, E, T, R, S, B, N, W(WRe5-26) |
| Humidity |
0 to 100% (Voltage 0V to 1V scaling conversion) |
| Filter |
Off, 2,5,10,20,40 Filter operation is on a moving average basis. The average value of the set sampling count is used. |
| Scan Speed |
0.1s/10 Ch Maximum |
| Trigger Functions |
Repeat Trigger |
Off, On |
| Trigger Types |
Start/Stop: Data capture starts/stops when a trigger is generated. |
| Trigger Conditions |
Start: Off, Level, External digital, Date
Stop: Off, Level, External digital, Date,Time |
| Alarm Judgement modes |
Analog: H, L, Window In, Window Out • Logic: H,L • Pulse: H, L, Window In, Window Out |
| Alarm Output |
Output Format |
Open Collector output (100 k Ohm pull-up resistance) |
| Output Conditions |
Level judgement, window judgement, logic pattern judgement, pulse judgement |
| Pulse/Logic Input |
Pulse input: 1ch/ Logic input:1ch |
| Pulse Input |
Revolutions mode |
Spans 50, 500, 5000, 50k, 500k, 5M, 50M 500M RPM |
| Counts Mode |
Spans 50, 500, 5000, 50k, 500k, 5M, 50M 500M Counts/sec |
| Inst. Mode |
Spans 50, 500, 5000, 50k, 500k, 5M, 50M 500M Counts/ec |
| Max number of pulse inputs |
Counts, Inst modes: 50k/sampling interval • Revs 50k/s |
| PC interface |
USB (Ver.1.1) |
| Internal memory devices |
Memory capacity |
• Internal memory: 3.5MB • USB memory: Depends on the type of memory used |
| Memory contents |
• Setup conditions • Measured data • Screen copy |
| Statistical calculation |
Types of operation |
Avarage value, peak value, maximum value, minimum value, RMS |
| Number of operations |
2 can be set simultaneously |
| Method |
Realtime operation
Note: When the Digital screen + Calculation Display screen has been specified, the calculation results are displayed. |
| Search functions |
Search the captured data for the required number of points |
| Scaling function |
4 points can be set for each channel |
| Display |
Size |
3.5'' TFT colour LCD (320 x 240 dots) |
| Display screen |
Waveform display |
Waveform screen + digital screen, Waveform screen |
| Digital display |
Waveform screen + digital screen, digital screen + Calculation Display screen |
| A/D Converter |
16bits (out of which 14 are internally acknowledged) |
| Maximum permissible input voltage |
Between +/- terminals: 60Vp-p • Between input terminals and casing: 60Vp-p |
| Input Impedance |
1M Ohm ±5% |
| Withstand Voltage |
Between each input channel and main unit chassis, and also between each CHs: 1minute at 350V p-p |
| Operating Enviroment |
0 to 40°C, 30 to 80% RH |
| Power Supply |
AC adapter |
100 to 240 VAC, 50/60 Hz |
| DC input |
8.5 to 24 VDC) |
| Battery pack(optional) |
7.4VDC (2200mAh) |
| Power consumption |
28 VA or lower (AC drive) |
External dimensions |
194 x 122 x 41 mm |
| Weight (approx.) |
480g (excluding the AC adapter and battery) |
midi Logger Software Specifications
| CPU |
Pentium 4 (1.7 GHz or better) |
| Memory |
Minimum of 512 MB (1GB recommended) |
| Compatible OS |
Windows 2000/xp |
| Functions |
Device control, real-time data capture, file format conversion |
| Device Setting Ranges |
Input setting, memory settings, alarm settings, trigger settings |
| Captured Data |
Real-time transfer of data to PC, transfer of data captured to internal device memory |
| Items Displayed |
Analog waveforms, logic waveforms, pulse waveforms, digital values |
| Display Modes |
Digital values, waveforms (Y-T, X-Y), meter view, report view |
| Monitoring Function |
E-mail is sent to the specified e-mail address(es) when an alarm is generated. |
| File format conversion |
Conversion of selected or all data into CSV file format |
| Export to direct Excel file |
Data is saved to an Excel file at the maximum 1-s sampling rate |
| Statistics/Log Display |
Displays maximum, minimum, and average values during measurement and generates a log of alarms issued, if any |
|