Agilent InfiniiVision 7000 Series Oscilloscopes: Engineered for the best signal visibility
The new Agilent InfiniiVision 7000 Series offers bandwidths up to 1 GHz and delivers an unparalleled deep memory waveform update rate of up to 100,000 waveforms per second. Each model, equipped with the industry’s largest display - a 12.1 inch XGA LCD display - comes in a package that is just 7" deep and weighs only 14 pounds.
Benefits of the Agilent InfiniiVision 7000 Series oscilloscopes include:
Biggest display Oscilloscopes are visual tools and large, high-resolution screens make the product better. Bigger displays have become increasingly important as general purpose scopes need more space to display digital and serial signals in addition to traditional scope channels. The increased display size helps you easily view up to 20 channels simultaneously with serial protocol. At 12.1 inches the display is nearly 40% bigger than any competitive model.
Fastest update rate InfiniiVision Series 7000 models provide the industry's fastest update rate - up to 100,000 waveforms per second - eliminating two common errors that can cause engineers to miss critical information: unresponsive controls with deep memory turned on, and traditional architectures that are blind to changes in signals being tested. For designs with both digital and analog signal measurements, the update rate is 5,000 times faster than any other available scope. This allows the user to view critical signal detail that competitive scopes miss with their longer dead times.
Turn knobs and the instrument responds instantly and effortlessly. Need to also view digital channels? The instrument stays responsive. Decoding serial packets? Offering the industry’s only hardware-accelerated serial bus decode, Agilent’s InfiniiVision series delivers serial debug without compromising analog measurements.
Most comprehensive suite of applications
Packed with numerous general purpose features, InfiniiVision 7000 Series oscilloscopes can be customized with a number of unique software options. Supported applications give engineers meaningful insight into application-specific problems. Applications include:
- Serial decode and trigger for I2C, SPI, CAN, LIN, FlexRay, RS-232 and other UART
- Rapid core-assisted debug of designs with Xilinx or Altera FPGAs
- Segmented Memory for analysis of laser pulses, radar bursts and serial packets.
- Power Measurement Application for a full suite of power measurements.
- Offline PC viewing and sharing of previously acquired scope data.
- View the Agilent InfiniiVision Introduction Demo
Agilent has been the fastest growing oscilloscope supplier since 1997 (source: Prime Data, 2007). Wonder why? Agilent engineers developed the InfiniiVision 7000 Series with advanced technology that will allow you to see more subtle signal detail and more infrequent events than any other scope on the market. See the InfiniiVision 7000 Series oscilloscope—the industry's best for signal viewing.
|
Software Applications
|
Offline viewing and analysis (Accessory - B4610A)
Need to view and analyze scope data away from your scope? Need to share measurement data with geographically dispersed team members? Save your scope data to a USB or network drive and import the data into a PC-based offl ine viewer. Pan and zoom. Use searching and fi ltering to gain insight on analog and digital buses. Email the data to team members who can use the same tool at their PCs.
Secure environment mode (Accessory - Option SEC)
Option SEC – secure environment mode provides the highest level of security by ensuring internal nonvolatile memory is clear of all setup and trace settings in compliance with National Industrial Security Program Operation Manual (NISPOM) Chapter 8 requirements. When this option is installed, it will store setup and trace settings to internal volatile memory only.
Volatile memory will be cleared during the power off cycle of the instrument. So you can move the instrument out of a secure area with confidence.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| DSO7054A |
500MHz, 4-channel oscilloscope |
| $13031.00 |  |
|
| DSO7032A |
350MHz 2-channel oscilloscope |
| $7381.00 |  |
|
| DSO7034A |
350MHz 4-channel oscilloscope |
| $9215.00 |  |
|
| DSO7052A |
500MHz, 2-channel oscilloscope |
| $10476.00 |  |
|
| DSO7104A |
1GHz, 4-channel oscilloscope |
| $16066.00 |  |
|
| Model |
Options |
List Price |
Sale Price |
| U1241A |
True-rms 10000 Count Handheld DMM |
| $201.00 |  |
|
| N2732A |
Rackmount kit for 7000 Series oscilloscopes |
| $552.00 |  |
|
| N2733A |
Soft Carrying Case for 7000 Series oscilloscope |
| $180.00 |  |
|
| N2735A/002 |
Please call or email us for pricing and availability. |
| N2735A/004 |
Please call or email us for pricing and availability. |
| N2736A/002 |
Please call or email us for pricing and availability. |
| N2736A/004 |
Please call or email us for pricing and availability. |
| N2737A |
DSO to MSO Upgrade Kit for DSO710XA Scopes |
| $4281.00 |  |
|
| N5406A/001 |
Please call or email us for pricing and availability. |
| N5406A/002 |
Please call or email us for pricing and availability. |
| N5423A |
I2C and SPI Serial Bus Decode, for 4-channel InfiniiVision DSOs/MSOs |
| $1354.00 |  |
|
| N5424A |
CAN and LIN Serial Decode, for 4-channel InfiniiVision DSOs/MSOs |
| $1354.00 |  |
|
| N5457A |
RS232/UART Serial Decode and Trigger |
| $1354.00 |  |
|
| N5432A |
FlexRay Solution, for 4-channel InfiniiVision 6000 and 7000 Series MSOs |
$3190.00 |
| $3158.10 |  |
|
| N5434A/001 |
Please call or email us for pricing and availability. |
| N5434A/002 |
Please call or email us for pricing and availability. |
| N5454A |
Segmented Memory |
| $1354.00 |  |
|
| Opt. SEC |
Please call or email us for pricing and availability. |
| B4610A/010 |
Please call or email us for pricing and availability. |
| B4610A/020 |
Please call or email us for pricing and availability. |
| U1881A/001 |
Please call or email us for pricing and availability. |
| U1881A/002 |
Please call or email us for pricing and availability. |
Series Comparison
| Model |
Bandwidth |
Sample Rate |
Memory Depth |
Scope Channels |
Digital Channels |
| DSO7032A |
350 MHz |
2 GSa/s |
8 Mpts |
2 |
|
| DSO7034A |
350 MHz |
2 GSa/s |
8 Mpts |
4 |
|
| MSO7032A |
350 MHz |
2 GSa/s |
8 Mpts |
2 |
16 |
| MSO7034A |
350 MHz |
2 GSa/s |
8 Mpts |
4 |
16 |
| DSO7052A |
500 MHz |
4 GSa/s |
8 Mpts |
2 |
|
| DSO7054A |
500 MHz |
4 GSa/s |
8 Mpts |
4 |
|
| MSO7052A |
500 MHz |
4 GSa/s |
8 Mpts |
2 |
16 |
| MSO7054A |
500 MHz |
4 GSa/s |
8 Mpts |
4 |
16 |
| DSO7104A |
1 GHz |
4 GSa/s |
8 Mpts |
4 |
|
| MSO7104A |
1 GHz |
4 GSa/s |
8 Mpts |
4 |
16 |
Specifications
| Acquisition - scope channels |
| Sample rate |
MSO/DSO703xA: 2 GSa/sec each channel MSO/DSO705xA/710xA: 4 GSa/sec half channel*, 2 GSa/sec each channel Equivalent-time sample rate: 400 GSa/s (when real-time mode is turned off) |
| Memory depth Standard |
2 channels/4 channels 8 Mpts/4 Mpts |
| Vertical resolution |
8 bits |
| Peak detection |
MSO/DSO703xA: 500-ps peak detect MSO/DSO705xA/710xA: 250-ps peak detect |
| Averaging |
Selectable from 2, 4, 8, 16, 32, 64 … to 65536 |
| High resolution mode |
Average mode with avg = 1 12 bits of resolution when ≥10 μs/div at 4 GSa/s or ≥20-μs/div at 2 GSa/s |
| Filter |
Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever is less) with vectors on and in real-time mode |
| Acquisition - digital channels (7000 Series MSO or MSO-upgraded 7000 Series DSO) |
| Sample rate |
2 GSa/sec one pod**, 1 GSa/sec each pod |
| Maximum input frequency |
250 MHz |
| Memory depth Standard |
One pod/both pods (with scope channels turned off) - 8 Mpts/4 Mpts One pod/both pods (with scope channels turned on) - 2.5 Mpts/ 1.25 Mpts |
| Vertical resolution |
1 bit |
| Glitch detection |
2 ns (min pulse width) |
* - Half channel is when one of channel 1 or 2 is turned on, and/or one of channel 3 or 4 is turned on.
** - A pod is a group of eight digital channels, either 0-7 or 8-15.
| Vertical system - scope channels |
| Scope channels |
MSO/DSO7xx2A: Ch 1 and 2 simultaneous acquisition MSO/DSO7xx4A: Ch 1, 2, 3 and 4 simultaneous acquisition |
| Bandwidth (–3 dB)* |
MSO/DSO703xA: DC to 350 MHz MSO/DSO705xA: DC to 500 MHz |
| MSO/DSO710xA |
DC to 1 GHz |
| AC coupled |
MSO/DSO703xA: 3.5 Hz to 350 MHz MSO/DSO705xA: 3.5 Hz to 500 MHz |
| MSO/DSO710xA |
3.5 Hz to 1 GHz |
| Calculated rise time (=0.35/bandwidth) |
MSO/DSO703xA: 1 nsec MSO/DSO705xA: 700 psec MSO/DSO710xA: 350 psec |
| Single-shot bandwidth |
MSO/DSO703xA: 350 MHz MSO/DSO705xA: 500 MHz MSO/DSO710xA: 1 GHz (in half-channel mode) |
| Range** |
MSO/DSO703xA and MSO/DSO705xA: 2 mV/div to 5 V/div (1 MΩ or 50 Ω) MSO/DSO710xA: 2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω) |
| Maximum input |
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk CAT II 100 Vrms, 400 Vpk With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk |
| Offset range |
±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div; ±75 V on ranges >200 mV/div |
| Dynamic range |
±8 div |
| Input impedance |
1 MΩ ± 1% || 14 pF or 50 Ω ± 1.5%, selectable Coupling AC, DC BW limit 25 MHz selectable |
| Channel-to-channel isolation |
DC to max bandwidth >40 dB |
| Standard probes |
10073C shipped standard for each scope channel (1165A probes optional) |
| Probe ID |
Auto probe sense and AutoProbe interface Agilent- and Tektronix-compatible passive probe sense |
| ESD tolerance |
±2 kV |
| Noise peak-to-peak |
MSO/DSO703xA: 3% full scale or 3 mV, whichever is greater MSO/DSO705xA: 3% full scale or 3.6 mV, whichever is greater MSO/DSO710xA: 3% full scale or 4 mV, whichever is greater |
| DC vertical gain accuracy** |
±2.0% full scale |
| DC vertical offset accuracy |
≤200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value >200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value |
| Single cursor accuracy** |
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)} Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset, accuracy = ±{2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)} = ± 4.785 mV |
| Dual cursor accuracy** |
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)} Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset, accuracy = ±{2.0% (80 mV) + 0.4% (80 mV)} = ±1.92 mV |
| Vertical system - digital channels (MSO or MSO-upgraded DSO) |
| Number of channels |
16 logic timing channels – labeled D15 - D0 |
| Threshold groupings |
Pod 1: D7 - D0 Pod 2: D15 - D8 |
| Threshold selections |
TTL, CMOS, ECL and user-definable (selectable by pod) |
| User-defined threshold range |
±8.0 V in 10 mV increments |
| Maximum input voltage |
±40 V peak CAT I; transient overvoltage 800 Vpk |
| Threshold accuracy* |
±(100 mV + 3% of threshold setting) |
| Input dynamic range |
±10 V about threshold |
| Minimum input voltage swing |
500 mV peak-to-peak |
| Input capacitance |
~8 pF with flying leads |
| Input resistance |
100 kO ±2% at probe tip |
| Channel-to-channel skew |
2 ns typical, 3 ns maximum |
| Horizontal |
| Range |
MSO/DSO703xA: 2 nsec/div to 50 sec/div MSO/DSO705xA: 1 nsec/div to 50 sec/div MSO/DSO710xA: 500 psec/div to 50 sec/div |
| Resolution |
2.5 ps |
| Time base accuracy |
15 ppm (±0.0015%) |
| Vernier |
1-2-5 increments when off, ~25 minor increments between major settings when on |
| Delay range |
Pre-trigger (negative delay): Greater of 1 screen width or 1 ms Post-trigger (positive delay): 1 s to 500 seconds |
| Analog delta-t accuracy |
Same channel: ±0.0015% reading ±0.1% screen width ±20 ps Channel-to-channel: ±0.0015% reading ±0.1% screen width ±40 ps Same channel example (MSO/DSO705xA): For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width), delta-t accuracy = ±{0.0015% (10 µs) + 0.1% (50 µs) + 20 ps} = 50.17 ns |
| Logic delta-t accuracy |
Same channel: ±0.005% reading ±0.1% screen width ±(1 logic sample period, 1 ns) Channel-to-channel: ±0.005% reading ±0.1% screen width ±(1 logic sample period) ±chan-to-chan skew Same channel example: For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width), delta-t accuracy = ±{0.005% (10 µs) + 0.1% (50 µs) + 1 ns} = 51.5 ns |
| Modes |
Main, zoom, roll, XY, segmented (optional) |
| XY |
Bandwidth: Max bandwidth Phase error @ 1 MHz: <0.5 degrees Z Blanking: 1.4 V blanks trace (use external trigger on MSO/DSO7xx2A, channel 4 on MSO/DSO7xx4A) |
| Reference positions |
Left, center, right |
| Segmented memory re-arm time |
10 µs maximum, 7.5 µs typical |
| Trigger system |
| Sources |
MSO7xx2A: Ch 1, 2, line, ext, D15 - D0 DSO7xx2A: Ch 1, 2, line, ext MSO7xx4A: Ch 1, 2, 3, 4, line, ext, D15 - D0 DSO7xx4A: Ch 1, 2, 3, 4, line, ext |
| Modes |
Auto, normal (triggered), single |
| Holdoff time |
~60 ns to 10 seconds |
| Trigger jitter |
15 ps rms |
| Selections |
Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, FlexRay, USB, I2C, SPI, Nth edge burst |
| Edge |
Trigger on a rising, falling, alternating or either edge of any source |
| Pattern |
Trigger at the beginning of a pattern of high, low, and don’t care levels and/or a rising or falling edge established across any of the analog and digital channels, but only after a pattern has stabilized for a minimum of 2 nsec. The scope channel’s high or low level is defined by that channel’s trigger level. The logic channel’s trigger level is defined by the threshold for the pod, 0 - 7 or 8 - 15. |
| Pulse width |
Trigger when a positive- or negative-going pulse is less than, greater than, or within a specified range on any of the source channels. Minimum pulse width setting: 5 ns (MSO/DSO703xA scope channels) 2 ns (MSO/DSO705xA/710xA scope channels) 2 ns (logic channels on 7000 Series MSO or MSO-upgraded 7000 Series DSO) Maximum pulse width setting: 10 s |
| TV |
Trigger using any scope channel on most analog progressive and interlaced video standards including HDTV/EDTV, NTSC, PAL, PAL-M or SECAM broadcast standards. Select either positive or negative sync pulse polarity. Modes supported include Field 1, Field 2, all fields, all lines, or any line within a field. TV trigger sensitivity: 0.5 division of sync signal. Trigger holdoff time can be adjusted in half field increments. |
| Sequence |
Arm on event A, trigger on event B (edge or pattern), with option to reset on event C or time delay. |
| CAN |
Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start of frame (SOF) bit (standard). N5424A option supports triggering on remote frame ID (RTR), data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors, acknowledge error and overload frame. |
| LIN |
Trigger on LIN (local interconnect network) sync break at beginning of message frame (standard). N5424A option supports triggering on frame ID. |
| FlexRay |
N5432A option supports trigger on FlexRay frame ID or time slot or specific error condition, along with cycle-base and repetition-cycle filtering. |
| USB |
Trigger on USB (universal serial bus) start of packet, end of packet, reset complete, enter suspend, or exit suspend on the differential USB data lines. USB low speed and full speed are supported. |
| I2C |
Trigger on I2C (inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data values. Also trigger on missing acknowledge, address with no acq, restart, EEPROM read, and 10-bit write. |
| SPI |
Trigger on SPI (serial protocol interface) data pattern during a specific framing period. Supports positive and negative Chip Select framing as well as clock idle framing and user-specified number of bits per frame. |
| RS-232/UART |
This application eliminates the need to manually decode bus traffic. Using data captured on the scope or digital channels, the application provides the ability to easily view the information sent over a RS-232 serial bus. Display real-time time-aligned decode of transmit and receive lines. The application also enables triggering on RS-232/UART conditions. |
| Duration |
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a value, greater than a time value with a timeout, or inside or outside of a set of time values. Minimum duration setting: 2 ns Maximum duration setting: 10 s |
| Nth edge burst |
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge count: 65,536. |
| Autoscale |
Finds and displays all active scope and logic (for 7000 Series MSO) channels, sets edge trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage >10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz. |
| Scope channel triggering |
| Range (internal) |
±6 div from center screen |
| Sensitivity* |
<10 mV/div: greater of 1 div or 5 mV; =10 mV/div: 0.6 div |
| Coupling |
AC, ~10 Hz on MSO/DSO703xA/705xA/710xA), DC, noise reject, HF reject and LF reject (~50 kHz) |
| Digital (D15 - D0) channel triggering (7000 Series MSO or MSO-upgraded 7000 Series DSO) |
| Threshold range |
(user defined) ±8.0 V in 10 mV increments |
| Threshold accuracy* |
±(100 mV + 3% of threshold setting) |
| Predefined thresholds |
TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V |
| External (EXT) triggering |
MSO/DSO7xx2A (2-/2+16-ch models) |
MSO/DSO7xx4A (4-/4+16-ch models) |
| Input impedance |
1 MΩ ± 3% || 14 pF or 50Ω |
2.14 kΩ ±5% |
| Maximum input |
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk 5 Vrms with 50-Ω input |
±15V |
| Range |
DC coupling: trigger level ±1 V and ±8 V |
±5 V |
| Sensitivity |
For ±1 V range setting DC to 100 MHz, 100 mV; For ±8 V range setting DC to 100 MHz, 250 mV; >100 MHz to bandwidth of oscilloscope: 500 mV |
DC to 500 MHz: 500 mV |
| Coupling |
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz) |
| Probe ID |
Auto probe sense and AutoProbe interface Agilent- and Tektronix-compatible passive probe sense |
| Display system |
| Display |
12.1-inch (255 mm x 184 mm) diagonal color TFT LCD |
| Throughput of scope channels |
Up to 100,000 waveforms/sec in real-time mode |
| Resolution |
XGA – 768 vertical by 1024 horizontal points (screen area); 640 vertical by 1000 horizontal points (waveform area) 256 levels of intensity scale |
| Controls |
Waveform intensity on front panel. Vectors on/off; infinite persistence on/off, 8 x 10 grid with intensity control |
| Built-in help system |
Key-specific help displayed by pressing and holding key or softkey of interest. Language support for 11 languages including English, German, French, Russian, Japanese, Traditional Chinese, Simplified Chinese, Korean, Spanish, Portuguese and Italian. |
| Real-time clock |
Time and date (user adjustable) |
| Measurement features |
| Automatic measurements |
Measurements are continuously updated. Cursors track last selected measurement. Up to four measurements can be displayed on screen at any one time. |
| Voltage (scope channels only) |
Peak-to-peak, maximum, minimum, average, amplitude, top, base, overshoot, preshoot, RMS, standard deviation (AC RMS) |
| Time |
Frequency, period, + width, – width and duty cycle on any channel. Rise time, fall time, X at max Y (time at max volts), X at min Y (time at min volts), delay, and phase on scope channels only. |
| Counter |
Built-in 5-digit frequency counter on any channel. Counts up to the scope’s bandwidth (1 GHz max). The counter resolution can be increased to 8 digits with an external 10-MHz reference. |
| Threshold definition |
Variable by percent and absolute value; 10%, 50%, 90% default for time measurements |
| Cursors |
Manually or automatically placed readout of horizontal (X, ?X, 1/?X) and vertical (Y, ?Y). Additionally logic or scope channels can be displayed as binary or hex values. |
| Waveform math |
One function of 1-2, 1x2, FFT, differentiate, integrate, square root. Source of FFT, differentiate, integrate: scope channels 1 or 2, 1-2, 1+2, 1x2. |
| FFT |
| Points |
Fixed at 1000 points |
| Source of FFT |
Scope channels 1 or 2 (or 3 or 4 on MSO/DSO7xx4A only), 1+2, 1-2, 1*2 |
| Window |
Rectangular, flattop, Hanning |
| Noise floor |
–50 to –90 dB depending on averaging |
| Amplitude |
Display in dBV, dBm at 50 O |
| Frequency resolution |
0.05/time per div |
| Maximum frequency |
50/time per div |
| Storage |
| Save/recall (non-volatile) |
10 setups and traces can be saved and recalled internally. Optional secure environment mode ensures setups and traces are stored to internal volatile memory so data is erased when power is removed. Compliant to NISPOM Chapter 8 requirements. |
| Storage type and format |
USB 1.1 host ports on front and rear panels Image formats: BMP (8-bit), BMP (24-bit), PNG (24-bit) Data formats: X and Y (time/voltage) values in CSV format, ASCII XY and binary format and .alb for offline viewing on a PC Trace/setup formats: Recalled |
| I/O |
| Standard ports |
USB 2.0 high-speed device, two USB 1.1 host ports, 10/100BaseT LAN, XGA video output |
| Max transfer rate |
USB (USBTMC-USB488): 3.5 Mbytes/sec 100 Mbps LAN (TCP/IP): 1 Mbytes/sec |
| Supported printers via USB |
For a list of currently supported printers visit www.agilent.com/fi nd/Infi niiVision-printers |
| General characteristics |
| Physical size |
(WxHxD) 17.9” x 11.7” x 8.6” (45.4 cm x 29.8 cm x 22 cm) with legs extended, with screen cover on 17.9” x 10.9” x 6.8” (45.4 cm x 27.7 cm x 17.3 cm) with legs contracted, with screen cover on |
| Weight; Net |
5.9 kg (13 lbs) Shipping: 9.3 kg (20.5 lbs) |
| Probe comp output |
Frequency ~2 kHz; Amplitude ~5 V |
| Trigger out |
When Triggers is selected (delay ~17 ns) 0 to 5 V into high impedance 0 to 2.5 V into 50 O When Source Frequency or Source Frequency/8 is selected 0 to 580 mV into high impedance 0 to 290 mV into 50 O Max frequency output: 350 MHz (in source frequency mode when terminated in 50 O) 125 MHz (in source frequency/8 mode when terminated in 50 O) |
| 10 MHz ref in/out |
TTL out, 180 mV to 1 V amplitude with 0 to 2 V offset |
| Kensington lock |
Connection on rear panel for security |
| Power requirements |
| Line voltage range |
~Line 120 W max, 96-144 V/48-440 Hz, 192-288 V/48-66 Hz, automatic selection |
| Line frequency |
50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC |
| Power usage |
110 W max |
| Environmental characteristics |
| Ambient temperature |
Operating –10 °C to +55 °C; non-operating –65 °C to +71 °C |
| Humidity |
Operating 95% RH at 40 °C for 24 hr; non-operating 90% RH at 65 °C for 24 hr |
| Altitude |
Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft) |
| Vibration |
Agilent class B1 and MIL-PRF-28800F; class 3 random |
| Shock |
Agilent class B1 and MIL-PRF-28800F; class 3 random; (operating 30g, 1/2 sine, 11 ms duration, 3 shocks/axis along major axis, total of 18 shocks) |
| Pollution degree2 |
Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation must be expected. |
| Typical operator noise |
30 dBa at front of instrument, 35 dBa at rear of instrument. |
| Indoor use |
Rated for indoor use only |
| Other |
| Measurement categories |
CAT I: Mains isolated CAT II: Line voltage in appliance and to wall outlet |
| Regulatory information |
Safety IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 1010.1:1992 UL 61010B-1:2003 |
| Supplementary information |
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly. The product was tested in a typical configuration with HP/Agilent test systems. |
* - Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** - 2 mV/div setting for 100 MHz models and 2 mV/div is a magnification of 4 mV/div setting for 350 MHz to 1 GHz models. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting and 32 mV for 2 mV/div sensitivity setting.
|