Agilent Modules and Accuracy Specifications

  Multiplexer Actuator Matrix RF Multiplexer Multifunction
  34901A 34902A1 34908A 34903A 34904A 34905A 34906A 34907A
General
Number of channels 20 + 2
2/4 wire
16
2/4 wire
40
1 wire
20
SPDT
4 x 8
2 wire
Dual 1 x 4 See page 22 of data sheet
50 Ω 75 Ω
Connects to Internal DMM          
Scanning Speed 60 ch/s 250 ch/s 60 ch/s          
Open/Close Speed 120/s 120/s 70/s 120/s 120/s 60/s  
Input
Voltage
(dc , ac rms)2
300V 300V 300V 300V 300V 42V  
Current
(dc , ac rms)
1A 50mA 1A 1A 1A 0.7A  
Power
(W , VA)
50W 2W 50W 50W 50W 20W  
DC characteristics
Offset Voltage 3 <3uV <6uV <3uV <3uV <3uV <6uV  
Initial Closed Channel R 3 <1Ω <1Ω <1Ω <0.2Ω <1Ω <0.5Ω  
Isolation ch-ch, ch-earth >10GΩ >10GΩ >10GΩ >10GΩ >10GΩ >1GΩ  
AC characteristics
Bandwidth4 10 MHz 10 MHz 10 MHz 10 MHz 10 MHz 2 GHz 5 2 GHz 5  
Insertion Loss (dB) 10 MHz -0.1 -0.1  
100 MHz -0.4 -0.4  
500 MHz -0.6 -0.5  
1 MHz -1 -1  
1.5 MHz -1.2 -1.5  
2 MHz -3 -2  
SWR 10 MHz 1.02 1.02  
100 MHz 1.05 1.05  
500 MHz 1.20 1.25  
1 MHz 1.20 1.40  
1.5 MHz 1.30 1.40  
2 MHz 1.40 2.0  
ch-ch Cross Talk (dB)4 10 MHz -45 -45 -186 -45 -33 -100 -85  
100 MHz -85 -75  
500MHz -65 -65  
1MHz -65 -65  
1.5MHz -45 -40  
2MHz -35 -35  
Risetime             <300ps  
Signal delay             <3ns  
Capacitance HI-LO <50pF <50pF <50pF <10pF <50pF <20pF  
LO-Earth &60;80pF &60;80pF <80pF <80pF <80pF  
Volt-Hertz limit   108 108 108 108 108 1010  
Other
T/C Cold Junction Accuracy3 (typical) 0.8ºC 0.8ºC 0.8ºC8          
Switch life No Load (typical) 100M 100M 100M 100M 100M 5M 5M  
Rated Load (typical)7 100K 100K 100K 100K 100K 100K 100K 100K
Temperature Operating all cards ¾ 0ºC to 55ºC
Storage all cards ¾ -20ºC to 70ºC
Humidity (non-condensing) all cards ¾ 40ºC / 80% RH
  1. Not recommended for connection to ac line without external transient suppression
  2. Channel-to-channel or channel-to-earth
  3. Errors included in DMM measurement accuracy specifications
  4. 50Ω source, 50Ω load
  5. Bandwidth direct to card SMB connectors
  6. Isolation within channel 1 to 20 or 21 to 40 banks is -40 dB
  7. Applies to resistive loads only
  8. Thermocouple measurements not recommended with 34908A module due to common lo configuration


  Accuracy specifications

± (% of reading + % of range)1
Includes measurement error, switching error, and transducer conversion error
  Range3 Frequency, etc. 24 Hour2
23&186;C ± 1&186;C
90 Day
23&186;C ± 5&186;C
1 Year
23&186;C ± 5&186;C
Temperature
Coefficient
0&186;C–18&186;C, 28&186;C–55&186;C
DC voltage
  100.0000mV   0.0030 + 0.0035 0.0040 + 0.0040 0.0050 + 0.0040 0.0005 + 0.0005
1.000000V   0.0020 + 0.0006 0.0030 + 0.0007 0.0040 + 0.0007 0.0005 + 0.0001
10.00000V   0.0015 + 0.0004 0.0020 + 0.0005 0.0035 + 0.0005 0.0005 + 0.0001
100.0000V   0.0020 + 0.0006 0.0035 + 0.0006 0.0045 + 0.0006 0.0005 + 0.0001
300.000V   0.0020 + 0.0020 0.0035 + 0.0030 0.0045 + 0.0030 0.0005 + 0.0003
True RMS AC Voltage4
  100.0000mV to 100.0000V 3 Hz–5 Hz 1.00 + 0.03 1.00 + 0.04 1.00 + 0.04 0.100 + 0.004
5 Hz–10 Hz 0.35 + 0.03 0.35 + 0.04 0.35 + 0.04 0.035 + 0.004
10 Hz–20 kHz 0.04 + 0.03 0.05 + 0.04 0.06 + 0.04 0.005 + 0.004
20 kHz–50 kHz 0.10 + 0.05 0.11 + 0.05 0.12 + 0.05 0.011 + 0.005
50 kHz–100 kHz 0.55 + 0.08 0.60 + 0.08 0.60 + 0.08 0.060 + 0.008
100 kHz–300 kHz5 4.00 + 0.50 4.00 + 0.50 4.00 + 0.50 0.20 + 0.02
  300.0000V 3 Hz–5 Hz 1.00 + 0.05 1.00 + 0.08 1.00 + 0.08 0.100 + 0.008
5 Hz–10 Hz 0.35 + 0.05 0.35 + 0.08 0.35 + 0.08 0.035 + 0.008
10 Hz–20 kHz 0.04 + 0.05 0.05 + 0.08 0.06 + 0.08 0.005 + 0.008
20 kHz–50 kHz 0.10 + 0.10 0.11 + 0.12 0.12 + 0.12 0.011 + 0.012
50 kHz–100 kHz 0.55 + 0.20 0.60 + 0.20 0.60 + 0.20 0.060 + 0.020
100 kHz–300 kHz5 4.00 + 1.25 4.00 + 1.25 4.00 + 1.25 0.20 + 0.05
Resistance6
  100.0000kΩ 1mA current source 0.0030 + 0.0035 0.008 + 0.004 0.010 + 0.004 0.0006 + 0.0005
1.000000kΩ 1mA 0.0020 + 0.0006 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
10.00000kΩ 100mA 0.0020 + 0.0005 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
100.0000kΩ 10mA 0.0020 + 0.0005 0.008 + 0.001 0.010 + 0.001 0.0006 + 0.0001
1.000000MΩ 5.0mA 0.002 + 0.001 0.008 + 0.001 0.010 + 0.001 0.0010 + 0.0002
10.00000MΩ 500nA 0.015 + 0.001 0.020 + 0.001 0.040 + 0.001 0.0030 + 0.0004
100.0000MΩ 500 nA/10MΩ 0.300 + 0.010 0.800 + 0.010 0.800 + 0.010 0.1500 + 0.0002
Frequency and Period7
  100mV to 300V 3 Hz–5 Hz 0.10 0.10 0.10 0.005
5 Hz–10 Hz 0.05 0.05 0.05 0.005
10 Hz–40 Hz 0.03 0.03 0.03 0.001
40 Hz–300 kHz 0.006 0.01 0.01 0.001
DC Current (34901A only)
  10.00000mA < 0.1V burden 0.005 + 0.010 0.030 + 0.020 0.050 + 0.020 0.002+ 0.0020
100.0000mA < 0.6V 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.002 + 0.0005
1.000000A < 2V 0.050 + 0.006 0.080 + 0.010 0.100 + 0.010 0.005 + 0.0010
True RMS AC Current (34901A only)
  10.00000mA and4 1.000000A 3 Hz–5 Hz 1.00 + 0.04 1.00 + 0.04 1.00 + 0.04 0.100 + 0.006
5 Hz–10 Hz 0.30 + 0.04 0.30 + 0.04 0.30 + 0.04 0.035 + 0.006
10 Hz–5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
100.0000 mA8 3 Hz–5 Hz 1.00 + 0.5 1.00 + 0.5 1.00 + 0.5 0.100 + 0.06
5 Hz–10 Hz 0.30 + 0.5 0.30 + 0.5 0.30 + 0.5 0.035 + 0.06
10 Hz–5 kHz 0.10 + 0.5 0.10 + 0.5 0.10 + 0.5 0.015 + 0.06
Temperature Type 1-Year Accuracy9   Extended Range 1-Year Accuracy9  
Thermocouple10 B 1100ºC to 1820ºC 1.2ºC 400ºC to 1100ºC 1.8ºC  
E -150ºC to 1000ºC 1.0ºC -200ºC to -150ºC 1.5ºC  
J -150ºC to 1200ºC 1.0ºC -210ºC to -150ºC 1.2ºC  
K -100ºC to 1200ºC 1.0ºC -200ºC to -100ºC 1.5ºC 0.03ºC
N -100ºC to 1300ºC 1.0ºC -200ºC to -100ºC 1.5ºC  
R 300ºC to 1760ºC 1.2ºC -50ºC to 300ºC 1.8ºC  
S 400ºC to 1760ºC 1.2ºC -50ºC to 400ºC 1.8ºC  
T -100ºC to 400ºC 1.0ºC -200ºC to -100ºC 1.5ºC  
RTD R0 from 49Ω to 2.1kΩ -200ºC to 600ºC 0.06ºC     0.003ºC
Thermistor 2.2k, 5k, 10k -80ºC to 150ºC 0.08ºC     0.002ºC

  1. Specifications are for 1 hr warm-up and 61/2 digits, Slow ac filter
  2. Relative to calibration standards
  3. 20% over range on all ranges except 300 Vdc and ac ranges and 1 Adc and ac current ranges
  4. For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50 kHz, add 0.1% of range additional error
  5. Typically 30% of reading error at 1 MHz, limited to 1 x 108 V Hz
  6. Specifications are for 4- wire ohms function or 2-wire ohms using Scaling to remove the offset. Without scaling, add 4 ? additional error in 2-wire Ohms function
  7. Input > 100 mV. For 10 mV inputs multiply % of reading error x 10
  8. Specified only for inputs >10 mA
  9. For total measurement accuracy, add temperature probe error
  10. Thermocouple specifications not guaranteed when 34970A module is present